Appendix A. CNR1 Performance and Measurements under Different Conditions
A-2
TABLE A-1. Typical output signals of CNR1 under different meteorological conditions.
Explanation can be found in the text.
Day
night
Cloudy
clear
+20 ºC
- 20 ºC
CG3
Up
CG3
low
CM3
up
CM3
low
Pt 100
sky T
ground
T
d cloud 20 0 0 0-500
0-150 20 20 20
d cloud -20 0 0 0-500
0-150 -20 -20 -20
d clear 20 -100* 0 0-1300
0-400 20 1* 20
d clear -20 -100* 0 0-1300
0-400
-20 -53* -20
n cloud 20 0 0 0 0 20 20 20
n cloud -20 0 0 0 0 -20 -20 -20
n clear 20
-100***
0 0** 0 20 1*** 20
n clear -20
-100***
0 0** 0 -20
-53***
-20
* Values may suffer from the so-called window heating offset; the sun heats
the pyrgeometer window causing a measurement error of + 25 Watts per square
meter (maximum).
** Values may suffer from negative Infrared offsets, caused by cooling off of
the CM3 dome by Far Infrared radiation. The maximum expected offset value
is 15 Watts per square meter.
*** Values may suffer from dew deposition. This causes the CG3-up values to
rise from -100 to 0 Watts per square meter.