Install & Operating Manual
CleanSeal V3
Local Aerosol Injection with IntegriPod
For aerosol requirements with respect to the airflow, please refer to ISO 14644-3
or other applicable standard. The IntegriPod insert in the upstream duct will allow to comply with or
exceed ISO 14644-3 particle distribution performance requirements.
On Integrity or custom housings, a local injection port is accessible in the filter chassis of the housing.
PLEASE NOTE:
Space inside the CleanSeal housing is restricted
and access to the injection port fitting can be
difficult for hookup and release of the injection
line, especially with MX or MG filters.
This can be addressed with the hookup fitting
mounted on a tube and a plastic rod for
operating the release button.
For superior ease of use, a hookup tool is available.
Hookup tool
1072227
➔
Please contact your Camfil representative
Using the Hookup-Tool
➔
The tool cannot be inserted the wrong way,
due to a safety key :
➔
The injection fitting is self-sealing and will
only lock and open, once the O-ring on the
tool's tip has sealed the connection
➔
Check condition of the O-ring before using
the HookUp-Tool, replace if necessary.
➔
2 spare O-rings are packed with the tool
For further spare O-rings use
O-ring DIN3770 - 7,5 x 1,8 NBR70
Release
Button