Command Syntax
CALibrate:COEFFicients:GAIN:VOLTage[:AC] <cr>
Description
Loads values measured with an external DVM for calibration of AC
voltage output and metering.
The first 3 values are Voltages measured at the output relay of the
power source for each phase.
The second set of 3 values are Voltages measured at the external
voltage sense points (at the load) for each phase.
Parameters
External reference voltage readings for each phase
Parameter Format
n/a
Example
n/a
Command Syntax
CALibrate:COEFFicients:GAIN:CURRent[:AC] <cr>
Description
Loads values measured with an external shunt and DVM for
calibration of AC current metering.
The first 3 values are currents measured at the output of the power
source for each phase.
Parameters
External reference current reading
Parameter Format
n/a
Example
n/a
Query Format
No
Returned Data Format
n/a
Query Example
n/a
Command Syntax
CALibrate:COEFFicients:GAIN:VOLTage:DC <cr>
Description
Loads values measured with an external DVM for calibration of DC
voltage output and metering.
The first 3 values are Voltages measured at the output relay of the
power source for each phase.
The second set of 3 values are Voltages measured at the EXTernal
Sense point (at the load) for each phase
Parameters
None
Parameter Format
n/a
Example
n/a
Command Syntax
CALibrate:COEFFicients:GAIN:CURRent:DC <cr>
Description
Loads values measured with an external shunt and DVM for
calibration of DC current metering.
The first 3 values are currents measured at the output of the power
source for each phase.
Parameters
None
Parameter Format
n/a
Example
n/a