User and Programming Manual - Rev N
11.4 Making Measurements
The source has the capability to return current, peak current, crest factor, true power,
apparent power, power factor and voltage measurements.
Voltage Measurement
To measure the RMS voltage, use:
OUTP 1
/* Close output relay if it was not closed already.
MEAS:VOLT?
Current Measurements
To measure the RMS current, use:
MEAS:CURR?
To measure the peak current, use:
MEAS:CURR:AMPL:RES
/* resets track and hold for peak curr. meas.
/* It may be need to turn the output on or program
/* as specific voltage at this time before taking a
/* peak current reading.
MEAS:CURR:AMPL:MAX?
/* retrieves peak current reading.
To measure the current crest factor, use:
MEAS:CURR:CRES?
Power Measurements
To measure the true power, use:
MEAS:POW:REAL?
To measure the apparent power, use:
MEAS:POW:APP?
To measure the power factor, use:
MEAS:POW:PFAC?
112 April
2007
2001RP
Содержание 2001RP
Страница 4: ...User and Programming Manual Rev N Electrical Safety Symbols Used in This Manual 2001RP April 2007 iii...
Страница 33: ...User and Programming Manual Rev N Figure 3 7 Test Setup 2001RP 24 April 2007 2001RP...
Страница 47: ......
Страница 48: ...User and Programming Manual Rev N Figure 5 1 AC Power System Block Diagram 2001RP April 2007 39...
Страница 60: ...User and Programming Manual Rev N Figure 6 3 Internal Adjustments 2001RP April 2007 51...
Страница 65: ...User and Programming Manual Rev N Figure 7 1 Assembly Location 56 April 2007 2001RP...
Страница 68: ...User and Programming Manual Rev N Figure 7 2 Replaceable Fuse Locations 2001RP April 2007 59...
Страница 70: ...User and Programming Manual Rev N Figure 8 1 Rear Panel View 2001RP April 2007 61...
Страница 90: ......