User and Programming Manual - Rev N
Frequency
The output frequency is controlled with the FREQuency command. To set the output
frequency to 50 Hz, use:
FREQuency 50
Turning on output voltage at a specified phase angle
The following command sequence will turn the output voltage on at a specified phase angle.
This may be required to measure worst case inrush current.
/* Set up sequence
VOLT 0
Set output voltage to zero
OUTP 1
Close output relay
OUTP:STAR:STAT 1
Enable start phase mode
OUTP:STAR:PHAS 90
Set start phase angle to 90 degrees
OUTP:DROP
Put AC source in infinite drop mode
/* Turn on sequence
VOLT 230
Turn on voltage to 230 V at 90 degrees
Current Limit
To set the RMS current limit use:
CURRent <n>
where <n> is the RMS current limit in amperes.
If the load attempts to draw more current than the programmed limit, The AC source is
programmed to turn off its output if the RMS current limit is reached.
Note:
The CURRent command is coupled with the VOLTage:RANGe.This means
that the maximum current limit that can be programmed at a given time
depends on the voltage range setting in which the unit is presently
operating.
2001RP April
2007
111
Содержание 2001RP
Страница 4: ...User and Programming Manual Rev N Electrical Safety Symbols Used in This Manual 2001RP April 2007 iii...
Страница 33: ...User and Programming Manual Rev N Figure 3 7 Test Setup 2001RP 24 April 2007 2001RP...
Страница 47: ......
Страница 48: ...User and Programming Manual Rev N Figure 5 1 AC Power System Block Diagram 2001RP April 2007 39...
Страница 60: ...User and Programming Manual Rev N Figure 6 3 Internal Adjustments 2001RP April 2007 51...
Страница 65: ...User and Programming Manual Rev N Figure 7 1 Assembly Location 56 April 2007 2001RP...
Страница 68: ...User and Programming Manual Rev N Figure 7 2 Replaceable Fuse Locations 2001RP April 2007 59...
Страница 70: ...User and Programming Manual Rev N Figure 8 1 Rear Panel View 2001RP April 2007 61...
Страница 90: ......