User and Programming Manual - Rev N
3.4 Mechanical
Installation
The power source has been designed for rack mounting in a standard 19 inch rack. The unit
must be supported from the sides with optional rack slides. See Accessory Equipment/Rack
Slides in Section 1.2. The cooling fan at the rear of the unit must be free of any obstructions
which would interfere with the flow of air. A 2.5 inch (6.35cm) clearance should be
maintained between the rear of the unit and the rear panel of the mounting cabinet. Keep in
mind that adequate ventilation of the mounting cabinet must be maintained for sufficient
cooling of the power source. Use louvered side panels, and/or perforated rear panels to
assure adequate airflow. Also, the air intake holes on the sides of the power source must not
be obstructed when mounted inside cabinet. See Figure 3-1.
The rack mount slide mounting holes centerlines on the side of the power source are 2.30
inches / 58.4 mm above the bottom edge of the front panel.
Note:
The 2001RP series models cannot be mounted in a cabinet by just
using the front panel rack ears. They require additional support.
Figure 3-5: Rack Mount Slides (-RMS option) position
3.5 Input
Wiring
The AC Power Source must be operated from a three-wire single-phase service. The mains
source must have a current rating greater than or equal to 35 A for the low input range. Use
the supplied safety cover/strain relief when connecting the input wires to the power source.
Refer to Figure 3-2 and Figure 3-6for the input power connections.
Also loop the input ground wire through the supplied ferrite filter (Ci P/N 250508) core 5
times (5 turns) for 230V input models (AWG#12) or 3 turns for 115V input models
(AWG#10). Enclose ferrite core with the safety cover or leave it outside the safety cover if
insufficient space is available.
When connecting the input wires to input terminal block, make sure the safety grounding
wire from the strain relief grommet including the 5 turns around the ferrite core is longer than
the Line and Neutral wire so that the ground wire will be the last one to take any strain. For
low-range input operation (115V) we recommend #10 AWG input wires as a minimum. For
high-range input operation (230V) #12AWG is recommended. For European applications
(230V), the following input cord set may also be used:
Panel Components # 86518030 Rated 16A, 250V, 70
°
C, 3x 1.5mm² conductor size. VDE
approved. Install as described above for discreet wiring.
2001RP April
2007
19
Содержание 2001RP
Страница 4: ...User and Programming Manual Rev N Electrical Safety Symbols Used in This Manual 2001RP April 2007 iii...
Страница 33: ...User and Programming Manual Rev N Figure 3 7 Test Setup 2001RP 24 April 2007 2001RP...
Страница 47: ......
Страница 48: ...User and Programming Manual Rev N Figure 5 1 AC Power System Block Diagram 2001RP April 2007 39...
Страница 60: ...User and Programming Manual Rev N Figure 6 3 Internal Adjustments 2001RP April 2007 51...
Страница 65: ...User and Programming Manual Rev N Figure 7 1 Assembly Location 56 April 2007 2001RP...
Страница 68: ...User and Programming Manual Rev N Figure 7 2 Replaceable Fuse Locations 2001RP April 2007 59...
Страница 70: ...User and Programming Manual Rev N Figure 8 1 Rear Panel View 2001RP April 2007 61...
Страница 90: ......