IT6860B User manual
用户使用手册
23
Trigger Mode
(
>TRIG
)
IT6860B series power supply supports two different trigger modes (MANU and BUS).
If select MANU, press
(Shift) +
Enter
(Trigger) can generate a trigger signal.
If select BUS, then sending command *TRIG can generate a trigger signal.
Factory default set is MANU.
Group set
(
MEM GROUP
)
IT6860B series power supply enables customer to store up to 9*8 groups of different
operating parameters in a non-volatile memory for quick recall.
How to understand 9*8
9*8
: There are 8 groups; each group includes nine different memory locations. The definition
of group is as follows:
GRP1: means to Save/Reall parameters in memory location from 1 to 9.
Operation: Press
(Shift) +
Recall
(Save) + numeric keys 1to 9(
Recall
+numeric keys 1
to 9)
GRP2: means to Save/Reall parameters in memory
location from 10 to 18.
GRP3-GRP8 is counted on the analogy of the same rule.
Timer setup
(
>TIMER SET
)
TIMER SET item is used to enable or disable timer function.
If select ON, means to enable output Timer. Output will turn off after value set in menu
“Timer set” counted down to zero. Meanwhile, VFD will appear a mark of “Timer”.
If select OFF, means to disable Timer function.
Factory default set is ON.
Reset to factory setting
(
>
RESET
)
RESET is used to return to the factory default setup.
If select YES, means to return to factory default setup.
If select NO, means to keep the current setup unchanged.
Factory default setup is as follows:
P-MEM=KEEP
P-OUT=OFF
COMM=RS232
CMD=SCPI
BEEP= ON
KNOB=ON