Personal Ambulatory Oxygen System
Provider Technical Manual
20771733 D 5
eQuinox Oxygen System Specifications
Dimensions (H x W X D)
13.6 x 10.6 x 7.4 (inches)
(34.5 x 26.9 x 18.8 cm)
Weight
eQuinox
Battery Pack
12.2 lbs (5.53 kg) unit only
14.0 lb (6.35 kg) w/ 12-cell battery installed
12 Cell Battery Pack - 1.85 lbs (0.84 kg)
24 Cell Battery Pack - 3.6 lbs (1.63 kg)
Flow Settings
Continuous Flow (measured in Liters Per Minute LPM)
Pulse Dose (measured in mL)
0.5 to 3.0 LPM (0.5 liter increments)
Settings 1.0-6.0: 16-96mL (8mL increments)
Settings 7-9: 128mL, 160mL, 192mL
Continuous Flow Accuracy
+/- 10% or 200ml/min, whichever is greater
Oxygen Concentration
90% (+5.5/-3%) for all flow settings
Maximum System Pressure
15 psig (103.5 kPa)
Oxygen Output Pressure
5.0 psig (34.5 kPa) nominal
Oxygen Concentration
Status Indicator
Green Light = Normal Operation
Yellow Light = Warning or Caution, less than 85% ± 3%
Red Light Flashing = Abnormal Operation, less than 70% ± 5%
Nominal Sound Level
3.0 LPM Continuous Flow
0.5 LPM Continuous Flow
46 dB(A)
40 dB(A)
Operating Environment
Temperature (Unit ONLY, without Battery Pack)
Humidity
32º F to 104º F (0° to 40°C)
10% to 95%, Non-condensing, 82.4°F (28°C) Maximum Dew point
Storage Environment
Temperature
Temperature (Battery Pack)
Humidity
-4º F to 140º F (-20° to 60°C)
-4°F to 122°F (-20° to 50°C)
Up to 95% Non-condensing
Altitude Operating Range
-1254 to 13,130 feet (-382 to 4,000 meters)
Nominal Power
0.5 LPM Continuous flow
3 LPM Continuous Flow
1.0 Pulse Dose Setting
6.0 Pulse Dose Setting
Battery Charging
60 Watts
110 Watts
45 Watts
95 Watts
Approx. 50 W depending on device setting
Nominal Battery Pack Operating Time
See Battery Duration Chart
Continuous Flow Indication
Expressed in liters per minute (LPM)
Audible Alarm Indicators
See Alarm Condition and Alarm Codes
Back-Up Alarm Power
9V Internal Battery
Filters
Air Inlet, HEPA, Compressor Intake