Personal Ambulatory Oxygen System
Provider Technical Manual
20771733 D 30
VI. Removing/Replacing the Product
Tank
1. Using a pair of small cutters, snip the end of the tie-wrap on the
soft tubing that connects the ATF to the Product Tank and discard
the tie-wrap.
Figure 46
2. Gently pull the soft tubing from the oxygen port on the ATF.
Figure 47
3. Using a pair of small cutters, snip the end of the tie-wrap on the
bottom of the clear rubber elbow that connects the Product Tank
to the aluminum tubing going to the flow tube on the Control
Board.
Figure 48
4. The Product Tank can now be removed by gently pulling upward.
To install a new Product Tank, repeat steps 1-3 in reverse.
Note:
Ensure to reinstall the tie-wrap on the soft tubing
connecting the ATF to the Product Tank as well as the
clear rubber elbow that connects to the aluminum
tubing going to the flow tube on the Control Board.