Personal Ambulatory Oxygen System
Provider Technical Manual
20771733 D 20
eQuinox Voice Message Alert Descriptions Table
eQuinox Voice Message Alert Descriptions
Condition
Standard eQuinox Voice Message
Message Repeat Interval
Flow rate error > +/- 10%
“Low output flow. Check the cannula.”
Continuous until condition is removed.
30 minutes runtime remaining on battery
“30 minutes remaining on battery.”
One instance
15 minutes runtime remaining on battery
“15 minutes remaining on battery.”
One instance
Low battery
“Low on battery. Please plug in external
power.”
Message played every 29 seconds until loss of
power state.
Change in continuous flow setting
“Continuous mode flow xx liter per minute.”
(xx = [0.5L, 1.0L, 1.5L, 2.0L, 2.5L, 3.0L])
One instance
Change in pulse flow setting
“Pulse dose mode xx milliliter.”
(xx = [16mL, 32mL, 48mL, 64mL, 80mL, 96mL,
128mL, 160mL, 192mL])
One instance
Patient’s breath rate not detected
“Cannot detect breathing.”
One instance
Loss of power to unit
“Loss of power. Please plug in external power
or battery.”
Continuously until condition is removed.
Unit is powered on in Pulse Mode and is using
AC power supply.
“The device is running. External power
connected. Pulse dose mode xx milliliter.”
(xx = [16mL, 32mL, 48mL, 64mL, 80mL, 96mL,
128mL, 160mL, 192mL])
After start-up beep, on instance.
Unit is powered on in Continuous Mode and
is using AC power supply.
“The device is running. External power
connected. Continuous dose mode xx liter
per minute.”
(xx = [0.5L, 1.0L, 1.5L, 2.0L, 2.5L, 3.0L])
After start-up beep, on instance.
Unit is powered on in Continuous Mode and
is using battery.
“The device is running. Using battery.
Continuous dose mode xx liter per minute.”
(xx = [0.5L, 1.0L, 1.5L, 2.0L, 2.5L, 3.0L])
After start-up beep, on instance.
Unit is powered on in Pulse Mode and is using
battery.
“The device is running. Using battery. Pulse
dose mode xx milliliter.”
(xx = [16mL, 32mL, 48mL, 64mL, 80mL, 96mL,
128mL, 160mL, 192mL])
After start-up beep, on instance.
While unit is running from battery, the AC
power supply is connected.
“External power connected.”
One instance
While unit is running from AC power with the
battery installed and the AC power supply is
removed.
“Using battery.”
One instance
When powering unit off.
“The device is now turned off.”
One instance
Table 8