12
Torque Minimum/Maximum
Set a minimum or maximum torque value.
From the Torque/Timer Menu Screen:
1. Press the TORQUE MINIMUM or TORQUE MAXIMUM
button to open the Keypad Screen.
2. Enter desired parameter. The set value is displayed on
the Torque/Timer Menu Screen, TORQUE MINIMUM
/TORQUE MAXIMUM button and Home Screen,
TORQUE Button.
3. Press PLAY Button. If the calculated max torque value is
less than the set value, the calculated max torque will be
displayed in red on the TORQUE Button.
Note:
If attempting to set a value outside of the allowable
limit, the default value will display. Minimum and maximum
values are always based on the actual torque value. When
setting a maximum torque, speed parameter must be set
first. Mixer will not operate if max torque is set to zero.
Note:
When the corresponding min or max value is reached,
the mixer will stop or continue operation as set (see Continue
After Alarm).
Note:
The calculated max torque value (red) represents the
maximum torque capability based on the speed.
T
orque Alarms
Set an audible alarm (5 short beeps) to sound when MIN/
MAX Torque is reached.
From the Torque / Timer Menu Screen:
1. Press TORQUE ALARMS button to toggle between
alarm on and alarm off.
2. Audible Alarm icon, indicating on/off setting appears
on the Torque/Timer Menu Screen, TORQUE ALARMS
button and Home Screen, TORQUE button.
Continue After Alarm
Set the mixer to continue/stop mixing when the set Torque
Min/Max is reached.
From the Torque / Timer Menu Screen:
1. Press CONTINUE AFTER ALARM button to toggle
between on and off.
2. When set to on, the mixer will continue mixing
indefinitely after the Torque Min/Max is reached. Set
Torque value will flash.
3. When set to off, the mixer will stop mixing when the
Torque Min/Max is reached.
4. Continue After Alarm icon, indicating on/off setting
appears on the Torque/Timer Menu Screen,
CONTINUE AFTER ALARM button and Home Screen,
TORQUE button.