V560 User Manual
13
3.4. SCALE INHIBIT
It is possible to inhibit the scales count in this way:
•
by sending a NIM signal through one of the two "VETO" connectors located on the front
panel.
•
by setting the VME VETO (access to the address Base +%52; see § 4.6).
The VME VETO is cleared:
•
by pushing the front panel push-button "MAN CLR";
•
by resetting the VME VETO (access to the address Base +%54; see § 4.6).
3.5. SCALE CLEAR
All the scales are cleared in the following cases:
•
by sending a NIM signal through one of the two "CLEAR" connectors located on the front
panel;
•
by pushing the front panel push-button "MAN CLR";
•
by accessing via VME the address Base + %50 (Scale Clear see § 4.7);
•
by generating the VME signal SYSRES.
3.6. CHANNELS TEST
If the module is configured with 16 independent channels it is possible to increment all channels:
•
by sending a NIM pulse through one of the two "TEST" connectors located on the front panel.
•
by accessing via VME the address Base + % 56 (Scale Increase see § 4.5).