5.3.17
“MST” Command
The ‘MST\r’ command returns the value of the power supply internal status
register (16 bit).
Replies from the LiAM power supply module to this command are in the
following format:
#MST:
value
\r
where:
value
is the ASCII representation of the internal status register value,
composed by 4 hexadecimal digits, and corresponding to the 16-bit wide status
register.
The MST command, being a reading command, returns a response in any
module condition.
Examples:
MST example with the module ON, ramping (and voltage rails at mid-level):
MST
\r
#MST:5001
\r