Safety and Installation
LiAM User’s Manual
24
2.10
Suggested cable sections
The suggested cable sections for the LiAM rear panel connections are all
included in the range 0.2 mm
2
→ 4 mm
2
, i.e. 24 AWG → 11 AWG, and presented in
Table 1
.
mm
2
AWG
AC Mains
1
17
Interlocks and Output Status
0.25
23
Output Terminals
1.5 → 2
15 → 14
Table 1:
suggested conductor sections
Please remember connect the power supply to the EARTH ground connection
first.
2.11
Voltage Rail Fuses
The voltages used for generating the power “rails” on the LiAM output stage
are protected with two fuses (one for each rail), placed on the rear panel.
The current rating of both these fuses is
8A-T
(standard 5 × 20 mm
dimension) and they are used to protect the output stages and circuitry from fault
conditions.
Figure 15:
fuse-holder
A simple screw-driver is needed in order to remove fuse holders and to replace
these fuses, as shown in
Figure 15
.