Problem
Solutions
1. MOBILE does not work after
pressing the ON/OFF switch
- Check the correct connection of AC/DC adapter to MOBILE and electrical mains
- If you’re using batteries, verify charge conditions and their correct position in the compartment
- For further information please see tables I and II
2. MOBILE does not nebulize or
nebulizes slowly
- Check the HI-FLO nebulizer. If it’s clogged, clean and disinfect as indicated in this manual. If problem
continues, replace nebulizer.
- Check that nozzle is in its correct position into the base of the nebulizer.
- If you’re using a oily medication, dilute it with physiological solution
3. Too Noisy
- Contact dealer or authorized technical service
If the unit does not nebulize once the above conditions have been checked, we suggest to contact an authorized technical service or to use the
warranty, if this one is still valid.
IN CASE OF ANOMALIES OR WRONG OPERATION CONTACT CA-MI TECHNICAL SERVICE BEFORE ANY INSPECTION IS
DONE.
AFTER CONTROL OF TECHNICAL SERVICE, CA-MI DOES NOT GIVE GUARANTEE TO NAY DEVICE THAT HAS BEEN
TAMPERED.
STANDARD ACCESSORIES
ACCESSORIES
HI-FLO KIT (Nebulizer HI-FLO, adult mask, pediatric mask, air tube, mouth-piece, nosepiece)
AC/DC Adapter
Power Cord for AC/DC adapter (for Adapter mod. UE20-120167SPA1)
Car Plug Cable
1 Spare Air Filter
Carrying Bag
1 – Air Tube
2 – Nebulizer Tank
3 – Nebulization Nozzle
4 – Nebulizer Top
5 – Mouthpiece
6 – Adult Mask
7 – Pediatric Mask
8 – Nosepiece
Only use the original accessories supplied by CA-MI with MOBILE.
Check the air filter every 25 hours of operation and replace it when turns dirty.
For replacement open the batteries cover on the bottom of device, and pull out (fig. 6) the air filter for replacement.
For each individual patient it’s recommended to use the HI-FLO nebulizer for 6 months or for a maximum of 120 treatments.
The HI-FLO nebulizer must be replaced after a long period of inactivity, if it is deformed or broken, or if the HI-FLO nebulizer nozzle is blocked by dry
medicine, dust, ecc..
Use the “nose piece” accessory only if expressly indicated by your doctor and paying attention
NEVER
to introduce inside the nose the nasal bifurcation,
but only bring it as close as possible.
Fig. 6 (Filter replacement)
DON’T USE MOBILE WITHOUT AIR FILTER