To insert batteries, position MOBILE turning it over a flat surface, with enough room. To access the batteries compartment, open the cover set on the
base (fig. 1 – 2), pulling the two levers on the base and raising the cover until complete removal.
FIG. 1
FIG. 2
The batteries shall be placed in the dedicated and NOT-REMOVABLE battery compartment positioned inside the unit.
Insert batteries in this compartment, one by one (fig. 5), carefully following these precautions:
-
Don’t use damaged batteries or batteries with unknown conditions;
-
Don’t insert used batteries, whose origin is unknown or with missing identification and make sure that batteries have technical specifications as
indicated in this manual;
-
Use only batteries stored in a dry and cool place and protected by atmospheric agents. Never use wet or overheated batteries;
-
Insert batteries with negative and positive poles correctly oriented (please see indications shown in battery compartment);
-
Use only rechargeable batteries Ni-MH type AA – 1.2V with capacity between 1600mAh and 2500 mAh;
-
The kit of n°8 batteries placed each time into MOBILE must have
SAME
characteristics and capacities:
NEVER MIX TOGETHER BATTERIES
WITH DIFFERENT CAPACITIES
(e.g: 4 batteries of 1600mAh with 4 batteries of 2300mAh). This kind of operation might seriously damage
batteries during MOBILE operation or during recharging cycle.
FIG. 3
Close the cover as shown in pictures n.4 e n.5.
Prepare the nebulizer HI-FLO with medication and accessories prescribed by doctor, connect it to MOBILE with the air tube, then press the ON/OFF
switch to start the treatment.
FIG. 4
FIG. 5