WARNING: BEFORE PROCEEDING, ALWAYS MAKE SURE NOT TO START RECHARGE CYCLE WITJ NOT-RECHARGEABLE 
ALKALINE BATTERIES. NEVER RECHARGE NON-RECHARGEABLE ALKALINE BATTERIES.

Insert batteries Ni-MH type AA – 1.2V in MOBILE, then connect it to the AC/DC adapter (or car plug cable) and to the electrical mains. 
To start recharge cycle, press the ON/OFF switch for at least 3 seconds. Rapid flashing of the green light signals the recharge cycle has started.
To increase batteries life, we recommend you follow instructions of batteries manufacturer and recharge them only when they are completely discharge.

During recharge cycle MOBILE cannot be operated. Pressing the ON/OFF switch will operate the compressor interrupting the 
recharging cycle. During preparation and performing of recharging cycle, luminous indications are shown as in the table II.

Recharge time indications and rechargeable batteries autonomy

It’s recommended to use only rechargeable batteries Ni-MH type AA – 1.2V with capacity between 1600mAh and 2500mAh.

The manufacturer cannot be considered responsible for problems caused by batteries with specifics different from those ones indicated here.

Recharge time of batteries with capacity 1600mAh:  

1.5 

 4 hours about

Batteries autonomy with capacity 1600mAh:  

40 minutes max.

Recharge time of batteries with capacity 2500mAh:  

 6.5 hours about

Batteries autonomy with capacity 2500mAh :  

90 minutes max.

- Don’t store away discharged batteries, because their autonomy or integrity could be reduced or damaged; 
- If batteries are not being used for a long time proceed with the recharge operations once every three months at least, to avoid 

damaging;

- Always keep batteries in a dry and cool place (15-20°C)

Operations by not-rechargeable alkaline batteries (not included):

In case of extreme necessity, this device can also be used with non-rechargeable alkaline batteries. To be able to use MOBILE with these batteries, you 
have to proceed as described in the previous chapter, paying extremely attention to the following points:

-

Insert non-rechargeable alkaline batteries type AA – 1.5V , making always sure THE RECHARGE CYCLE NEVER STARTS, when these batteries 
are located inside. 

NEVER PLACE NON-RECHARGEABLE ALKALINE BATTERIES INSIDE MOBILE, MIXING THEM WITH RECHARGEABLE ONES

.

NEVER RECHARGE NON-RECHARGEABLE ALKALINE BATTERIES

. We recommende to pull out non-rechargeable alkaline batteries at the end of

each treatment, to avoid the accidental start of recharge cycle.

Non observance of these warnings might result in serious danger situations for which the manufacturer for which the manufacturer cannot be held liable.
We therefore recommended limit the use non-rechargeable alkaline batteries fully charged non-rechargeable alkaline batteries can guarantee operation
of MOBILE for 20 minutes approx.

TABLE I – LUMINOUS INDICATIONS DURING THE OPERATION OF MOBILE

Indication of Light

State of the Unit

Problem / Cause

Solution

Red Flashing

Operating

Discharged battery

If batteries are rechargeable, start recharge cycle. If batteries are 
non-rechargeable alkaline, replace them

Red Fixed

Blocked

High temperature of the motor

Wait for the cool down until yhe red light turns off

Red Fixed for 3 
seconds

Blocked

Overloading – Short Circuit

Contact technical service

No light

Blocked

Defective AC/DC adapter or 
technical internal problem

Contact technical service

No light

Blocked

Inversions of poles

Disconnect the connector of AC/DC adapter,check the correct 
position of batteries, then re-connect and start MOBILE. If problem
continues, contact technical service.

www.nebulisercentre.co.uk

Содержание Mobile

Страница 1: ...PORTABLE COMPRESSOR NEBULIZER INSTRUCTION MANUAL www nebulisercentre co uk...

Страница 2: ...environments which have anaesthetic mixtures inflammable with air oxygen or nitric oxide Don t touch MOBILE with wet hands and always prevent the appliance is not touched by liquids The use of this d...

Страница 3: ...PPLY 100 240V 50 60Hz with power supplier in equipment model FUHUA UE20 120167SPA1 or UE24WCP 120167SPA or with car plug cable provided with unit or 12V dc with rechargeable batteries NI MH typeAA 1 2...

Страница 4: ...st IEC EN 61000 4 4 2kV power supply Mains power quality should be that of a typical commercial environments or hospital Surge IEC EN 61000 4 5 1kV differential mode Mains power quality should be that...

Страница 5: ...manual Before proceeding to the cleaning and maintenance operations and at the end of each use remember to turn off the switch and disconnect MOBILE from electrical power supply WARNING When MOBILE op...

Страница 6: ...in a dry and cool place and protected by atmospheric agents Never use wet or overheated batteries Insert batteries with negative and positive poles correctly oriented please see indications shown in...

Страница 7: ...15 20 C Operations by not rechargeable alkaline batteries not included In case of extreme necessity this device can also be used with non rechargeable alkaline batteries To be able to use MOBILE with...

Страница 8: ...mponent of the nebulizer except air tube removing medication residual and possible impurities Clean all parts in warm water Rinse thoroughly making sure that all deposits are washed away and let dry D...

Страница 9: ...E TO NAY DEVICE THAT HAS BEEN TAMPERED STANDARD ACCESSORIES ACCESSORIES HI FLO KIT Nebulizer HI FLO adult mask pediatric mask air tube mouth piece nosepiece AC DC Adapter Power Cord for AC DC adapter...

Отзывы: