Class II Insulation Device
CE marking in conformity with EC directive 93/42/CEE and subsequent changes
Warning: read instruction manual
Store in cool and dry place
Storing temperature: - 25 ÷ 70° C
Type B Device
~
Alternate current
Direct current
Hz
Mains frequency
ON / OFF
A copy of EC declaration may be requested form CA-MI S.r.l. – Via Ugo La Malfa, nr.31 – 43010 Pilastro (PR) Italy
Guidance and manufacturer’s declaration – Electromagnetic Emissions
The
MOBILE
aerosol is intended for use in the electromagnetic environment specified below.
The customers or the user of the
MOBILE
aerosol should assure that it’s used in such an environments.
Emission Test
Compliance
Electromagnetic environment - guidance
Irradiated / Conducted emission
CISPR11
Group 1
The MOBILE aerosol only used RF energy only for its internal
functioning. Therefore its RF emissions are very low and are not cause
interference in proximity of any Electronic appliances.
Irradiated / Conducted emissions
CISPR11
Class [B]
The MOBILE aerosol can be used in all environments, including
domestic and those connected directly to the public mains distribution
that supplies power to environments used for domestic scopes.
Harmonic emissions IEC/
EN 61000-3-2
Class [A]
Voltage fluctuations / flicker
emissions IEC/EN 61000-3-3
Complies
Immunity Test
Compliance
Electromagnetic environments - giudance
Electrostatic discharge (ESD)
IEC/EN 61000-4-2
± 6kV on contact
± 8kV in air
Floors should be wood, conceret or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient / burst
IEC/EN 61000-4-4
± 2kV power supply
Mains power quality should be that of a typical commercial
environments or hospital.
Surge IEC/EN 61000-4-5
± 1kV differential mode
Mains power quality should be that of a typical commercial
environments or hospital.
Loss of voltage, brief voltage
interruptions and variations
IEC/EN 61000-4-11
5%U
T
for 0.5 cycle
40%U
T
for 05 cycle
70%U
T
for 25 cycle
<5%U
T
for 5 sec
Mains power quality should be that of a typical commercial
environments or hospital. If the user of the MOBILE aerosol request
that the appliance operates continuosly, the use of a continuity unit is
recommended.
Magnetic field
IEC/EN 61000-4-8
3A/m
The power frequency magnetic field should be measured in the
intended installation location to assure that it’s sufficiently low.
Conducted Immunity
IEC/EN 61000-4-6
3Vrms 150kHz to 80MHz
-
Irradiated Conducted
IEC/EN 61000-4-3
3V/m 80MHz to 2.5 GHz
-
Nota U
T
is the value of the power supply voltage
DISPOSAL OF WASTE BATTERIES (Directive 2006/66/EC)
This symbol on the battery or on the packaging indicates that the battery provided with this product shall not be treated as household waste. By ensuring these
batteries are disposed of correctly, you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by
inappropriate waste handling of the battery. The recycling of the materials will help to conserve natural resources. At the end of their life hand the
batteries over to the applicable collection points for the recycling of waste batteries. For more detailed information about recycling of this product or battery, please
contact your local Civic Office, your household waste disposal service or the shop where you purchased the product.
www.nebulisercentre.co.uk