background image

Class II Insulation Device

CE marking in conformity with EC directive 93/42/CEE and subsequent changes

Warning: read instruction manual

Store in cool and dry place

Storing temperature: - 25 ÷ 70° C

Type B Device

~

Alternate current

Direct current

Hz

Mains frequency

ON / OFF

A copy of EC declaration may be requested form CA-MI S.r.l. – Via Ugo La Malfa, nr.31 – 43010 Pilastro (PR) Italy

Guidance and manufacturer’s declaration – Electromagnetic Emissions

The 

MOBILE 

aerosol is intended for use in the electromagnetic environment specified below.

The customers or the user of the 

MOBILE 

aerosol should assure that it’s used in such an environments.

Emission Test

Compliance

Electromagnetic environment - guidance

Irradiated / Conducted emission 
CISPR11

Group 1

The MOBILE aerosol only used RF energy only for its internal 
functioning. Therefore its RF emissions are very low and are not cause
interference in proximity of any Electronic appliances.

Irradiated / Conducted emissions 
CISPR11

Class [B]

The MOBILE aerosol can be used in all environments, including 
domestic and those connected directly to the public mains distribution 
that supplies power to environments used for domestic scopes.

Harmonic emissions IEC/
EN 61000-3-2

Class [A]

Voltage fluctuations / flicker 
emissions IEC/EN 61000-3-3

Complies

Immunity Test

Compliance

Electromagnetic environments - giudance

Electrostatic discharge (ESD)
IEC/EN 61000-4-2

± 6kV on contact

± 8kV in air

Floors should be wood, conceret or ceramic tile. If floors are covered 
with synthetic material, the relative humidity should be at least 30%.

Electrical fast transient / burst
IEC/EN 61000-4-4

± 2kV power supply

Mains power quality should be that of a typical commercial 
environments or hospital.

Surge  IEC/EN 61000-4-5

± 1kV differential mode

Mains power quality should be that of a typical commercial 
environments or hospital.

Loss of voltage, brief voltage 
interruptions and variations
IEC/EN 61000-4-11

5%U

T

 for 0.5 cycle

40%U

T

 for 05 cycle

70%U

T

 for 25 cycle

<5%U

T

 for 5 sec

Mains power quality should be that of a typical commercial 
environments or hospital. If the user of the MOBILE aerosol request 
that the appliance operates continuosly, the use of a continuity unit is 
recommended.

Magnetic field
IEC/EN 61000-4-8

3A/m

The power frequency magnetic field should be measured in the 
intended installation location to assure that it’s sufficiently low.

Conducted Immunity
IEC/EN 61000-4-6

3Vrms 150kHz to 80MHz 

-

Irradiated Conducted
IEC/EN 61000-4-3

3V/m 80MHz to 2.5 GHz

-

Nota U

T

 is the value of the power supply voltage

DISPOSAL OF WASTE BATTERIES (Directive 2006/66/EC)

This symbol on the battery or on the packaging indicates that the battery provided with this product shall not be treated as household waste. By ensuring these

batteries are disposed of correctly, you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by

inappropriate waste handling of the battery. The recycling of the materials will help to conserve natural resources. At the end of their life hand the

batteries over to the applicable collection points for the recycling of waste batteries. For more detailed information about recycling of this product or battery, please

contact your local Civic Office, your household waste disposal service or the shop where you purchased the product.

www.nebulisercentre.co.uk

Содержание Mobile

Страница 1: ...PORTABLE COMPRESSOR NEBULIZER INSTRUCTION MANUAL www nebulisercentre co uk...

Страница 2: ...environments which have anaesthetic mixtures inflammable with air oxygen or nitric oxide Don t touch MOBILE with wet hands and always prevent the appliance is not touched by liquids The use of this d...

Страница 3: ...PPLY 100 240V 50 60Hz with power supplier in equipment model FUHUA UE20 120167SPA1 or UE24WCP 120167SPA or with car plug cable provided with unit or 12V dc with rechargeable batteries NI MH typeAA 1 2...

Страница 4: ...st IEC EN 61000 4 4 2kV power supply Mains power quality should be that of a typical commercial environments or hospital Surge IEC EN 61000 4 5 1kV differential mode Mains power quality should be that...

Страница 5: ...manual Before proceeding to the cleaning and maintenance operations and at the end of each use remember to turn off the switch and disconnect MOBILE from electrical power supply WARNING When MOBILE op...

Страница 6: ...in a dry and cool place and protected by atmospheric agents Never use wet or overheated batteries Insert batteries with negative and positive poles correctly oriented please see indications shown in...

Страница 7: ...15 20 C Operations by not rechargeable alkaline batteries not included In case of extreme necessity this device can also be used with non rechargeable alkaline batteries To be able to use MOBILE with...

Страница 8: ...mponent of the nebulizer except air tube removing medication residual and possible impurities Clean all parts in warm water Rinse thoroughly making sure that all deposits are washed away and let dry D...

Страница 9: ...E TO NAY DEVICE THAT HAS BEEN TAMPERED STANDARD ACCESSORIES ACCESSORIES HI FLO KIT Nebulizer HI FLO adult mask pediatric mask air tube mouth piece nosepiece AC DC Adapter Power Cord for AC DC adapter...

Отзывы: