TRANS CAL
Model 7280
Page 34
Given :
Load cell model 8524-6050
strain gauge 350
Ω
Range: 50
kN
Output
signal
: 1.5016
mV/V
Zero output without fitting parts:
0.0020 mV/V
After selection of e.g. „SENSOR 1“ and after push of
ENTER
you get into the mode the sensor
can be applied in Adjust measuring range and decimal point (POINTDIS), UNIT and name of
the sensor (DESIGN).
POINTDIS
By pressing
▲
or
▼
you get to the following screens:
UNIT
ADJUST
DESIGN
After pressing SENSOR 1 and
ENTER
you get the following screen e.g.
10.00 Nm
Now in this example the number e.g. 50.00 can be set by pressing of
▲
or
▼.
If the beam is
below the count this can be changed by pressing of
▲
or
▼
. The changed count is taken by
press of
ENTER
.
After pressing 4 times
ENTER
you can set the decimal point by pressing
▲
oder
▼.
If the de-
sired decimal point reached you can taken this by press of
ENTER
.
Now return to the following screen:
POINTDIS
By pressing
▲
you get to the next screen:
UNIT