Model 7280
TRANS CAL
Page 21
UNIT
Entry of the physical unit, e.g. Kg, Ncm, t, kN, Nm, bar ...
DESIG
Name of the sensor, e.g. 1, 2, 3, DR 2112, silo, tank, mixer, scale1,
motor, test1...
ADJUST
Sensor type selection:
ACTIVE
without control signal, for transmitter 0-10 V
CURRENT for transmitter 4-20 mA.
PASSIVE
without control signal, for strain gauge (load) sensors
4.5 Adjustment procedure
ACT nCON
Active sensor without 100 % control signal
with following adjustment possibilities:
- adjust 0 % load and 100 % load
- adjust 0 % load and enter hub
(100 % load in V - 0 % load in V)
- enter 0 % load in V and the hub
(100 % load in V - 0 % load in V)
Select between 0 % load or nominal value by pressing
▲
or
▼
0 % LOAD unload sensor
or
NOMVALUE input of the nominal value in V
Select between load or nominal value by pressing
▲
or
▼
100 % LOAD Adjustment by 100 % load (apply nominal load)
or
NOMVALUE input of the nominal value in V
SAVE
Query for takeover of the adjusted data
PAS nCON Passive Sensor without 100 % control signal
with following adjustment possibilities:
- adjust 0 % load and 100 % load
- adjust 0 % load and enter 100 % load in mV/V
- enter 0 % load in mV/V and 100 % load in mV/V
Select between load or nominal value by pressing
▲
or
▼
0 % LOAD unload sensor.
or
NOMVALUE input of the nominal value in mV/V
Select between 100 % load or nominal value by pressing
▲
or
▼
100 % LOAD adjustment by load (apply nominal load).
or
NOMVALUE input of the nominal in mV/V
SAVE Query for takeover of the adjusted data