TRANS CAL
Model 7280
Page 26
5.2 Read-out of current sensor parameters
Sensor designation
8 Byte
ASCII
Final displayed value
2 Byte
compressed BCD-figure
Unit
3 Byte
ASCII
Sensor type and digit
1 Byte
0xAB: A ... Sensor type, B ... digit (binary coded)
Sensor type:
0xXXXX XXXX
||||
0000 ... active
with 100% 100% control signal adjust 0% load
and 100% load
0001 ... active
without 100% 100% control signal adjust 0% load and 100% load
0010 ... active
without 100% 100% control signal adjust 0% load and 100% load V
0011 ... active
without 100% 100% control signal adjust 0% load V and 100% load V
0100 ... passive
with 100% 100% control signal adjust 0% load and 100% load
0101 ... passive without 100% 100% control signal adjust 0% load and 100% load
0110 ... passive without 100% 100% control signal adjust 0% load V and 100% load mV
0111 ... passive without 100% 100% control signal adjust 0% load mV and 100% load mV/V
1000 ... current adjust 0% load
and 100% load
1001 ... current adjust 0% load
and 100% load mA
1010 ... current adjust 0% load mA and 100% load mA
Digit:
0xXXXX XXXX
||||
|000 ... _5000___
|001 ... _5,000__
|010 ... _50,00__
|011 ... _500,0__
|100 ... 5,000___
0%load
2 Byte
HEX value (MSB/LSB)
100% load
2 Byte
HEX value (MSB/LSB)
5.3 Read-out complete status
Status
2 Byte
general error condition of the 7280 measuring amplifier