TRANS CAL
Model 7280
Page 22
Adjustment procedure
CURRENT
Sensor with 4 ... 20 mA
with following adjustment possibilities:
- adjust 0 % load and 100 % load
- adjust 0 % load and 100 % load in mA
- enter 0 % load in mA and 100 % load in mA
Select between 0 % load or nominal value by pressing
▲
or
▼
0 % LOAD
unload sensor
or
NOMVALUE input of the nominal value in mA (fixed value 4-20 mA)
Select between 100 % load or nominal value by pressing
▲
or
▼
100 % LOAD adjustment by 100 % load (apply nominal load / nominal torque)
or
NOMVALUE input of the nominal value in mA
SAVE
Query for takeover of the adjusted data
If a sensor cannot be selected and/or does not change to the measuring
mode, there was a false adjustment for this sensor / parameter set.
→
Re-
adjustment