![Burkert 8205 Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/burkert-0/8205/8205_user-manual_2832363008.webp)
E-8-
8205
pH CONTROLLER 8205
2 SPECIFICATION
2.5 Technical Data 8205 pH Controller
2.5.1 pH controller type 8205 compact version specification
Ambient temperature
0 to 60
°
C (32 to 140
°
F)
Storing temperature
0 to 60
°
C (32 to 140
°
F)
Relative humidity
max 80 %
Enclosure
IP 65
Measuring range
0...14 pH
Measuring error
+/-0,2%, depending on electrode calibration
Temperature compensation
automatic (integrated Pt1000 or user parametred)
reference temperature 25
o
C (77
°
F)
Power supply:
12...30 VDC
Output signal
4...20 mA programmable ; proportional to the pH or temperature
Load
max. 1300
Ω
at 30 V ; max. 1000
Ω
at 24 V ; max. 550
Ω
at 15 V
Display
15 x 60 mm LCD 8 digits, alphanumeric,
15 segments, 9 mm high
Relay alarm output
1 relay, 3 A, 220 V, freely adjustable
Pulse Outputs
2 frequency and time duration adjustable outputs
- Relay Pulse Output:
3 A / 250 V max ; F < 1 Hz
- Transistor Pulse Output:
300 mA / 30 V max ; F < 17 Hz
- Triac Pulse Output:
1 A / 250 V max ; F < 17 Hz
Pt1000
SS 1.4571 (Ti 316)
Control Mode
P ; PI or PID programmable
Sensor housing
PVDF ; O-rings FPM/EPDM
Electronics housing
PC ; front plate polyester
Technical Data: Electrodes
GLS
STE
Housing
glass shaft
glass shaft
Fluid pressure
0-6 bar
0-3 bar
(0-87 psi)
(0-44 psi)
Fluid temperature
0-90
°
C
0-130
°
C
(32-194
°
F)
(32-266
°
F)
Max. pressure at max. temperature
4 bar (58 psi)
2 bar (29 psi)
Diaphragm
zirkondioxide
zirkondioxide
Reference electrolyte
gel
gel
LEI
SCH
HOL
Housing
glass shaft
glass shaft
Fluid pressure
0-2 bar
0-2 bar
0-6 bar
(0-29 psi)
(0-29 psi)
(0-87 psi)
Fluid temperature
0-60
°
C
0-40
°
C
0-90
°
C
(32-140
°
F)
(32-104
°
F)
(32-194
°
F)
Max. pressure at max. temperature
2 bar (29 psi)
2 bar (29 psi)
4 bar (58 psi)
Diaphragm
3 x zirkondioxide
none
none
Reference electrolyte
KCl 3-Molar
polymerised
polymerised