pH CONTROLLER 8205
4 OPERATION
Display selection and
increasing key
(numeric values)
impulses or automatic.
Choice of digit value
Steps from 0 to 9
PH
=
7.25
Relay Alarm: contact closed
Relay 2: contact closed
(Acid)
Acceptance of chosen
parameter or adjusted
value
4.1 Controller Operating and Control Elements
Direction downwards
in menu or sideways
for digit selection
Relay 1: contact closed
(Base)
The operation of the pH controller is classified according to 3 levels.
Main Menu
pH, temperature, output current, setpoint and working rates are displayed in the normal
function mode.
The "
HOLD
" function and electrode calibration function ("
PH CALIB
") can be accessed in this
menu. (see § 4.2)
Calibration Menu
The calibration mode allows adjustments of language, units, 4...20 mA output, output
frequencies and impulse durations, pH setpoint adjustment, regulation principle (P; PI;
PID), alarm thresholds, manual display selection, temperature compensation mode, and
filter selection. (see § 4.3).
Test Menu
The test menu allows the basic setting of the controller (Offset, Span, temperature
adjustment).
A pH or temperature value can be simulated via this menu, allowing the process to be
tested in the "dry condition".
The display of the instantaneous electrode voltage is available. (see § 4.4)