pH CONTROLLER 8205
4 OPERATION
4.4.4 pH Simulation
A pH (or T
°
) value can be simulated in this menu, allowing the user to test his system
without any liquid. The simulated value influences the current and pulse outputs and the
alarm relay.
Quit the sub-menu SIMUL by pressing
or
The simulation remains active until the user enters another sub-menu.
4.4.5 Display of electrode voltage
Display of the instantaneous value of the electrode voltage.
5 MAINTENANCE
5.1 Replacement of the pH electrode (compact controller only)
pH-electrodes have a limited service life, depending upon many parameters, such as the
chemical composition of the handled fluid, temperature, pressure, etc.
The electrode must be replaced if it shows visible damage (broken glass, fractures, etc.)
or if the messages "
WARNING
" or "
ERROR
" are displayed at the end of calibration.