- 46 -
ADDENDUM - 1011C DATA OUT VOLTAGES AND CONVERSIONS
Data Out Conntector Pins:
A
-DPI Dew point indication – binary output of 0 V or 5 V.
If 1011C is balanced and controlling, output is 5 V, else output is 0 V. Note: this is listed
as RTD in the manual.
C-
Balance Vbal (0-10 V range)
Output of signal from optics. 1011C is balanced and controlling on dew or frost when Vbal
is close to 5V.
E
-PRESSURE for Pressure sensor option. Vpress (0-10V range)
Press(mb) = 100* (Vpress – 1.6)/.64
G
-TEMPERATURE (Dew Point) Vdf (0-10V range), Tdf range (-75 to +50)
Tdf = 12.5 * Vdf - 75
B,D,F,H
are all corresponding ground connections for the outputs listed above.
Содержание 1011C
Страница 7: ...7 C B A Figure 1 Dew Point Hygrometer System shown with optional cables...
Страница 9: ...9...
Страница 12: ...12 Figure 3 Dew Point Sensing Unit A B C...
Страница 14: ...14 Figure 4 Control Indicator Unit F A B C D E...
Страница 17: ...17 Figure 6 Power Unit Internal...
Страница 23: ...23 Figure 9 Flow Test Fixture Hookup...
Страница 28: ...28 FIGURE 10 System Response Characteristics...
Страница 31: ...31 Recommended cable Belden YQ27540 Figure 11 Wiring Diagrams...
Страница 32: ...32 Recommended cable Belden YQ27540 Wiring Diagrams Continued...
Страница 33: ...33...
Страница 34: ...34 Figure 13 Vapor Pressure e vs Dew Point TD or Saturation Vapor Pressure es vs Temperature T...
Страница 36: ...36 Figure 15...
Страница 37: ...37 Figure 16 Aspirating Fixture Hookup...
Страница 38: ...38 Figure 17 1011C Hygrometer Complete...
Страница 40: ...40 Table 2 Dew Point to Frost Point Conversions...
Страница 41: ...41 Table 3 Dew Point Error due to Pressure Change in Sampling System...
Страница 47: ...47...
Страница 48: ...48...