8 | Cleaning and servicing
BÜCHI Labortechnik AG
40/56
Operation Manual SyncorePlus
8.2
Locking and unlocking the rotation
Locking the rotation
u
Press the navigation control 2 seconds.
ð
The rotation is locked.
ð
The status box shows the status locked.
Unlocking the rotation
Precondition:
R
The status box shows the status locked.
u
Press the navigation control 2 seconds.
ð
The rotation is unlocked.
8.3
Moving and retrieving the heating plate into a pre-defined position
For adjusting the unbalance compensation, the heating plate has to be in a pre-de-
fined position.
Moving the heating plate into a pre-defined position
u
Press the button
STOP
5 seconds.
ð
The heating plate moves to a pre-defined position.
ð
The heating plate is locked.
Retrieving the heating plate from a pre-defined position
u
Press the button
STOP
.
ð
The heating plate is unlocked.
8.4
Adjusting the unbalance compensation
The unbalance compensation is used to change the distance of a balancing weight
from the horizontal axis of the vortex movement.
A sensor monitors the vibration and controls the rotation speed.
NOTE
The default setting of the unbalance compensation is 23 mm. This default setting
covers all standardized operations.
u
Move the heating plate to the pre-defined
stopping position. See Chapter 8.3 "Moving
and retrieving the heating plate into a pre-de-
fined position", page 40
u
Insert the tool.