BÜCHI Labortechnik AG
Cleaning and servicing | 8
Operation Manual SyncorePlus
39/56
8
Cleaning and servicing
NOTE
Users may only carry out the servicing and cleaning operations described in this sec-
tion.
Any servicing and repair work which involves opening up the casing may only be car-
ried out by BUCHI service technicians.
u
Use only genuine BUCHI consumables and spare parts in order to ensure correct
operation of the device and preserve the warranty.
8.1
Regular maintenance work
CAUTION
Hot surfaces.
Skin burns from hot surfaces.
u
Let the instrument cool down sufficiently before carrying out any maintenance
work.
NOTE
It is the responsibility of the operators to select the correct cleaning agents and mate-
rials.
4
Do not use abrasive cleaning material.
4
Make sure that the cleaning agent meets the requirements of the safety data
sheets for the chemicals used.
4
Make sure that the cleaning agent is compatible with the materials used. See
Chapter 3.5 "Technical data", page 17
4
If you have any further questions, contact BUCHI Customer Service.
NOTE
Lock the instrument before carrying out any regular maintenance work. See Chap-
ter 8.2 "Locking and unlocking the rotation", page 40
Component
Action
Frequency
Casing
u
Wipe down the casing with a damp
cloth.
u
If heavily soiled, use ethanol or a mild
detergent.
Weekly
Warning symbols
u
Check that the warning symbols on
the instrument are legible.
u
If they are dirty, clean them.
Weekly
Rack mount
u
Wipe down the rack mount with a
damp cloth.
u
If heavily soiled, use ethanol or a mild
detergent.
Monthly
Base plate mount
u
Wipe down the base plate with a
damp cloth.
u
If heavily soiled, use ethanol or a mild
detergent.
Monthly