background image

20

BÜCHI 

Rotavapor

 R-200/205

4.14 Combination with Büchi Vacuum Controller

V-800/805

The new Büchi Vacuum Controllers V-800/805 are optimally
adapted to the design of the Rotavapor R-200/205.

Scoe of delivery

If the Vacuum Controller V-800/805 and valve unit are ordered
together with the Rotavapor R-200/205, these are already
installed into the units.

Retrofitting

If a Rotavapor R-200 or R-205 is retrofitted, you must proceed
as follows:

Fixation

• Screw out screw 

!

 and pull the cover plate to the back.

This cover 

%

 will not be used any more.

• Remove the cable socket 

"

 and plug it into the Vacuum

Controller.

• Place the Vacuum Controller V-800/805 on the special

cover 

#

 and draw the cover from the rear side on the

Rotavapor. Fix the cover with the screw.

• Control cable 

$

 

for the Rotavapor

The holding device for the valve unit can be fixed on the left
side of the tower of the Rotavapor. The valve unit can be fixed
on the holding device.

Cabeling of the Rotavapor

&

Connect mains supply bath to the mains

'

Connect mains supply of rotary evaporator to the mains

(

R socket on Vacuum Controller is connected to the socket
on the Rotavapor

)

Switchbox  socket  on  the  V-500  is  connected  to  the
”PUMP” socket on the Vacuum Controller

*

Valve socket on Vacuum Controller is connected to the
valve unit 

,

+

CW  socket  on  Vacuum  Controller  is  connected  to  the
cooling water valve

,

Valve Unit

Picture 19: Installation of Vacuum Controller on Rotavapor

4 Putting into operation

!

#

"

$

*

+

)

'

(

&

,

%

Содержание Rotavapor R-200

Страница 1: ...e of Delivery 2 2 Safety 4 3 Function 9 4 Putting into operation 12 4 1 Point of use 12 4 2 Electrical connections 12 4 3 Heating bath 13 4 4 Steam tube and sealing system 13 4 5 Glass assembly 14 4 6...

Страница 2: ...d 40602 Glass assembly C uncoated or 40640 Plastic Glas coated 40642 Glass assembly S uncoated or 40650 Plastic Glas coated 40652 Glass assembly E uncoated or 40660 Plastic Glas coated 40662 Glass ass...

Страница 3: ...230 V 40738 B chi Vacuum Controller V 805 120 V 40739 Valve unit to V 800 805 37968 Holder for valve unit 37968 to 40810 Rotavapor R 200 205 Remote control for V 800 805 40735 Automatic sensor to V 80...

Страница 4: ...turbances reduced economic efficiency and product losses can result Warning Surface temperature more than 60 C 2 2 Requirements placed on operators This apparatus can only be used by expert personnel...

Страница 5: ...ly with the technical data are considered improper The operator assumes sole risk for any damages attributed to such improper use Especially the following uses are not permitted Use of the apparatus i...

Страница 6: ...by authorised repair personnel The apparatus cannot be put into operation with damaged glassware 2 6 Safety measures Regional and local laws and regulations must be respected Basically personal prote...

Страница 7: ...ator is responsible for training his personnel For this purpose you can be order these operating instructions in other languages These operating instructions must be available for operating personnel...

Страница 8: ...ctronic over temperature protection and a power up time limit The driving unit is equipped with an electronic over temperature protection A safety spring protects the vapour duct from dropping out Gla...

Страница 9: ...istilling solvents with low boiling points S with vertical condenser and reflux capability CR with cold traps and reflux capability for distilling solvents with low boiling points E with descending co...

Страница 10: ...tion also results in even mixing of the sample and thus prevents stationary over heating in the flask Maximum weight 3 kg Rotation drive The drive unit makes sure that the evaporating flask rotates ev...

Страница 11: ...h sealing system Electronics head Evaporating flask Glass assembly Receiving flask Combi clip Heating bath Display module for rotation speed and vapour temperature optional B chi Vacuum Controller opt...

Страница 12: ...aporator only in running hood If for space related reasons this is not possible the optional protective shield should be mounted and the exhaust gas should be led from the pump into the running hood 4...

Страница 13: ...he bath is switched on less often This measure helps to save energy up to 50 At the same time approximately 70 less water is used see Chapter 8 11 Miscellaneous You should turn on the heating bath aro...

Страница 14: ...cured with the clip foreseen for this purpose Only glassware in perfect condition is to be used The glassware cannot have tears stars or other damages You must check glassware for damages before each...

Страница 15: ...an tilt to the left which can result in glass breakage Hold the glass assembly when you loosen the handle screw 4 9 End catch when submerging the evaporating flask By adjusting the catch screw the low...

Страница 16: ...he rope has to be adjusted This can be done by the customer 1 If the servo jack can not be lowered to the end turn the screws 1 on the front side of the rotavapor 1 4 turn clockwise 2 Repeat step 1 if...

Страница 17: ...tubes as short as possible If no B chi Vacuum Controller is used a Woulff bottle is to be connected between the vacuum source and Rotavapor If a B chi Vacuum Controller is used a valve unit should be...

Страница 18: ...Picture 17 Tube schema Cooling Water Vacuum connection to valve unit Vacuum sensor to valve unit Connection to glass assembly Connection to pump Connection to Vacuum Controller Connection for inert ga...

Страница 19: ...to the back Lift the lower spring loaded catch of the front plate and remove the front plate Connect the display module with the flat belt cable in the apparatus Connect the 2 pole cable with the dis...

Страница 20: ...Controller V 800 805 on the special cover and draw the cover from the rear side on the Rotavapor Fix the cover with the screw Control cable for the Rotavapor The holding device for the valve unit can...

Страница 21: ...a the Vacuum Controller V 800 805 rotation switches off and the evaporating flask is raised from the bath Valve unit When the set point on the Vacuum Controller is reached the valve unit shuts off the...

Страница 22: ...200 205 Socket for vapour temperature sensor only for R 205 Control cable of Vacuum Controller on R 200 or R 205 Mains supply for V 800 805 5 2 Controls B chi Rotavapor R 200 Main switch On Off Settin...

Страница 23: ...jack the evaporating flask can be lowered into and raised from the water bath This operation requires no physical effort Motion is effected by an integrated motor The button for the corresponding dir...

Страница 24: ...DESIRED temperature can now be set After approximately 2 seconds the display jumps back to the ACTUAL value The temperature scale via the SET point potentiometer is 2 coloured Blue marks the water ra...

Страница 25: ...Allow cooling water to flow through the condenser at approximately 40 50 l h and not warmer than 20 C Fill solution into the evaporating flask The solution can also be sucked in using vacuum Start eva...

Страница 26: ...f back evapo rating Then continue distillation Repeat this process until all desired solvent is distilled off At the end of distillation start rotation pull up the flask and vent the system Turn off t...

Страница 27: ...ing point of the solvent is 40 C You can get the corresponding pressure from the Solvent Table next page Advantages associated with 60 C bath temperature The evaporating flask can be replaced without...

Страница 28: ...resulting in increase of distillation capacity or 2 increasing bath temperature Increase of T1 resulting in increase of distillation capacity By increasing the bath temperature not all of the additio...

Страница 29: ...695 118 1 049 44 Ethanol C2H6O 46 0 879 79 0 789 175 Ethylacetate C4H8O2 88 1 394 77 0 900 240 Heptane C7H16 100 2 373 98 0 684 120 Hexane C6H14 86 2 368 69 0 660 335 Isopropylalcohol C3H8O 60 1 699 8...

Страница 30: ...ystem has been installed Check sealing system incorrectly Seal is defect Replace seals Frequent switching of valve or pump System is untight Control all sealing points tubes and their connections Hyst...

Страница 31: ...eplacing with suitable new tubes Sealing system Visually check seal KD 34 and if it is strongly worn replace it with a new seal Do not grease seal abrasion and fat have abrasive effect Cleaning by rin...

Страница 32: ...with non abrasive cleaning agents for example bathroom cleaners If calcification is per sistent you can use acetic acid to remove it Then you must thoroughly rinse the bath Clean the housing with a m...

Страница 33: ...ting to the maximum catch clockwise The motor is to turn without interruption in each position of the regulator 6 4 Customer service Only authorised service personnel can perform work on or in the app...

Страница 34: ...on Clean the apparatus thoroughly Residues of chemicals must be removed completely and glass components must be cleaned Keep and transport the apparatus in its original packaging 7 2 Disposal So that...

Страница 35: ...tion check to make sure that it operates correctly according to Chapter 6 2 Manufacturing according to this Manual is prohibited Copyright remains at B chi AG 8 1 Condenser fixation Name Order number...

Страница 36: ...36 B CHI Rotavapor R 200 205 Picture 29 Explosion illustration glass assembly A 8 Spare parts and accessories...

Страница 37: ...connection 40133 Joint adapter with combi clip 40615 NS 29 2 32 complete PTFE drain disk 40625 Stop cocks NS 18 8 38 40627 Glass assembly A complete with 1 l 40630 receiving flask 03275 40627 00643 4...

Страница 38: ...38 B CHI Rotavapor R 200 205 8 Spare parts and accessories Picture 30 Explosion illustration glass assembly V...

Страница 39: ...nge screwed connection 40133 Glass assembly V complete with 1 l 40600 receiving flask 40601 40604 03275 40627 00646 40625 28096 V condenser 23875 23900 25435 40601 Tube nipple GL 14 compl Joint adapte...

Страница 40: ...40 B CHI Rotavapor R 200 205 8 Spare parts and accessories Picture 31 Explosion illustration glass assembly C...

Страница 41: ...oint adapter with combi clip 40615 NS 29 2 32 complete Stop cocks NS 18 8 38 40627 Stop cocks NS 18 8 38 40628 Glass assembly C complete with 1 l 40640 receiving flasks 03275 00646 04625 40628 28090 4...

Страница 42: ...42 B CHI Rotavapor R 200 205 8 Spare parts and accessories Picture 32 Explosion illustration glass assembly S...

Страница 43: ...adapter with combi clip 40615 NS 29 2 32 complete PTFE disk drain disk 40625 Shut off tap 40626 Stop cocks NS 18 8 38 40627 Glass assembly S complete with 1l receiving 40650 flasks without vacuum con...

Страница 44: ...44 B CHI Rotavapor R 200 205 8 Spare parts and accessories Picture 33 Explosion illustration glass assembly E...

Страница 45: ...with steam tube seal 40010 condenser side Teflon ring evaporator side Set flange screwed connection 40133 Rod 40609 Joint adapter with combi clip 40615 NS 29 2 32 complete PTFE disk drain disk 40625...

Страница 46: ...46 B CHI Rotavapor R 200 205 8 Spare parts and accessories Picture 34 Explosion illustration glass assembly CR...

Страница 47: ...ng evaporator side Set flange screwed connection 40133 Joint adapter with combi clip 40615 NS 29 2 32 complete PTFE disk drain disk 40625 Shut off valve 40626 Stop cocks NS 18 8 38 40627 Stop cocks NS...

Страница 48: ...nt adapter complete with combi clip 40617 To 29 2 42 Joint Joint adapter complete with combi clip 40618 To 34 5 35 Joint Vapour duct 40610 Set adapter seal complete 2 units 40030 Seal KD 34 complete 4...

Страница 49: ...58 08739 1000 ml 00435 00440 08762 2000 ml 00436 08765 08769 3000 ml 00437 08767 08770 SJ Evaporating flask pear shaped PLASTIC GLAS Content SJ 29 2 32 SJ 24 40 SJ 29 2 42 50 ml 33405 100 ml 33404 250...

Страница 50: ...use of the plastic glass coating is according to manufacturers instructions in water up to 60 possible The cleaning of the glass parts should be done by hand Receiving flask with ball joint 35 20 Con...

Страница 51: ...lation spider Name Order number Flanged evaporating flask 100 ml 40675 250 ml 40676 500 ml 40677 1000 ml 40678 2000 ml 40679 3000 ml 40680 Flanged drying flask 500 ml 40682 1000 ml 40683 2000 ml 40684...

Страница 52: ...ml Beaker flask For working with high viscosity and solid substances The large opening allows for smooth draining and cleaning Foaming solutions can be better processed in a beaker flask Beaker Drying...

Страница 53: ...e obtained for 200 ml or 500 ml Extraction unit complete 200 ml 40699 11744 11705 11585 11904 19051 00982 08560 Extraction unit complete 500 ml 40699 11745 11705 11388 11904 19051 00982 00989 Reducing...

Страница 54: ...ion and thus increases its service life Woulff bottle complete 25519 25518 32885 Woulff bottle 25518 Tube nipple GL 14 bend cpl 4 units 37287 Picture 47 Floating balls in waterbath 8 Spare parts and a...

Страница 55: ...c Water jet pump 02913 8 Spare parts and accessories Water control valve Water control valve complete 11606 Picture 50 Water control valve Vacuum gauge unit Vacuum gauge 18826 Y piece 10 mm 00394 Pict...

Страница 56: ...55 Photograph of valve unit Valve unit This valve unit with integrated separator is a necessary accessory if vacuum is controlled with a vacuum controller It interrupts the connection between the pum...

Страница 57: ...ion II II II overvoltage category IP class Contamination degree 2 2 2 Rotation speed range 20 280 rpm 20 280 rpm Maximum size of flask 3 It max 3 kg 3 It max 3 kg 3 It max 3 kg Temperature range 20 C...

Страница 58: ...frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is lik...

Страница 59: ...l scientific and medical high frequency instruments EN 61000 3 2 1995 1996 Limits for harmonic current emissions EN 61000 3 3 1995 Limitation of voltage fluctuations and flicker Canadian Radio Interfe...

Страница 60: ...60 B CHI Rotavapor R 200 205 9 Appendix...