BÜCHI Labortechnik AG
Appendix | 10
Operation Manual Pure Chromatography Instrument C-805
67/70
u
Adjust the following values if necessary:
Column
Flow rate
Speed / Purity
u
Tap the button
[Accept]
.
ð
The gradient is saved for a run.
ð
The dialog box closes.
10.3.3
LC-C18
The LC-C18 Tool uses isocratic HPLC separation data to provide a recommended
gradient for reversed phase separations of 2 components.
Specification of the samples needed:
Two isocratic runs on a HPLC column
Different mobile phase solvent concentrations
Retention times (t1 and t2) of the components from the chromatograms.
Precondition:
R
The display shows the dialog
Navigator
. See Chapter 10.3.1 "Open the naviga-
u
Select the
LC-C18
tab.
u
Select the HPLC column used from the drop-down list.
u
Select the weaker solvent at drop-down list A.
u
Select the stronger solvent at drop-down list B.
u
Enter the solvent concentrations used in the HPLC separations.
u
Enter the retention times for each separation under each chromatograph.
u
Select the column you wish to use from the drop-down list
ð
The column is selected.
ð
The display shows the default flow rate.