BÜCHI Labortechnik AG
Operation | 6
Operation Manual Pure Chromatography Instrument C-805
45/70
6.4.5
Selecting values on the dialog box Sample Loading
The display shows the dialog box
Sample Loading
during the starting phase of a
separation.
The following settings are possible:
Action
Explanation
[Lower flow rate for
specified time after in-
jection]
Reduce flow rate for a specified period, when the pressure
increases significantly at the start of the run after sample
injection.
[Flush 2nd solvent at
the end of the run]
Automatically cleaning the system and column.
[Override Start Vial]
Select the start vial according to your needs.
6.5
Performing a Flash separation using a method
6.5.1
Preparing the instrument
Time required:
approx. 30 sec.
Precondition:
R
All commissioning operations have been completed. See Chapter 5 "Installation",
page 18
u
Switch the
On/Off
master switch to On.
u
Tap the
On/Off
switch.
ð
The instrument is starting up.
6.5.2
Starting Flash separation using a method
Precondition:
R
The instrument is prepared. See Chapter 6.5.1 "Preparing the instrument",
page 45
R
R
The sample is prepared.
R
The waste bottle is empty.
u
Open the protection shield.
u
u
Close the protection shield.
u
Open an existing method. See Chapter 6.4.4 "Selecting an existing method",
page 44
u
Tap the button
[OK]
.
u
Tap the button
[Start]
on the
Run
panel.
u
According the requirements adjust the file name.
u
Tap the button
[OK]
.
ð
The display shows the dialog box
Sample Loading
.
u
According the requirements adjust the settings. See Chapter 6.4.5 "Selecting val-
ues on the dialog box Sample Loading", page 45