Büchi Labortechnik AG
Installation | 5
Operation Manual UniversalExtractor E-800
31/88
NOTICE
Risk of property damage and diminished performance due to use of unsuitable
power cables.
The power supply cables supplied with the product by BUCHI precisely match the
requirements of the device. If other power cables that do not meet those
requirements are used, the device may be damaged and/or its performance
diminished.
u
Use only the power supply cables supplied with the product or ordered
separately from BUCHI.
u
If using any other power supply cables, make sure that they match the
specifications on the type plate.
u
Make sure that all connected devices are grounded.
u
Make sure that the power plug is freely accessible at all times.
u
Insert the power cable into the connection labeled Power IN on the back of the
instrument.
u
Insert the power plug into the power socket.
5.4
Securing against earthquakes
The instrument has an earthquake fixing point to protect the device against falling.
1
Fig. 10: Securing against earthquakes
1
Lashing mount
u
Tie the lashing mount to a fixed point using strong cord or a wire.
5.5
Connecting the coolant hose
There are two possibilities for coolant supply:
Cooling water valve
BUCHI Recirculating Chiller