8 | Help with faults
BÜCHI Labortechnik AG
30/38
Operation Manual Inert Loop B-295 SE
u
Remove the connected oxygen sensor.
STOP
START
u
Connect the power cable.
u
Press the button
MENU
.
ð
The display shows
[PASS]
.
MENU
u
Press the following sequence of buttons with-
out interruption.
u
Press the button
MENU
.
ð
The display shows
[E:1]
MENU
u
Press the button
ENTER
.
u
Press the button
INCREMENT
or
DECRE-
MENT
to set calibration gas level (Air: 20.9 %
O
2
)
or
u
Press button
ENTER
to span the sensor.
ð
The display shows
[ - - - -]
.
u
Press the button
MENU
.
ð
The span value will be displayed on exit
as % of the initial calibration value.
ð
The calibration is passed when the value
is between 20 and 21.
ð
The calibration is finished.
MENU