7 | Cleaning and servicing
BÜCHI Labortechnik AG
26/38
Operation Manual Inert Loop B-295 SE
u
Close the condensate drain valve.
u
Remove the oxygen sensor. See Chap-
ter 8.2.1 "Removing the oxygen sensor",
page 31
u
Insert the screw plug.
u
Disconnect the process gas tubes at the con-
nected spray dryer.
u
Place the end of the process gas tubes in a
top position.
u
Place the end of the exhaust gas tube in a top
position.
u
Fill in about 2 liter cleaning liquid (e.g. ethanol)
into the process gas tube connected to the
connection marked
IN
until the liquid reaches
the exhaust connection.
u
Wait 5 minutes.