6 | Operation
BÜCHI Labortechnik AG
24/38
Operation Manual Inert Loop B-295 SE
u
Check the temperature.
6.4
Shutting down the instrument
Precondition:
R
The spray drying process is finished. See operation manual of the connected
spray dryer.
u
Switch off the spray gas. See operation manual of the connected spray dryer.
ð
The signal lamp oxygen is on.
u
Switch off temperature. See operation manual of the connected spray dryer.
u
Wait until the temperature is less than 70 °C.
u
Switch off the aspirator. See operation manual of the connected spray dryer.
ð
The signal lamp pressure is on.
u
Empty the solvent receiving vessel. See Chapter 7.3 "Empty the solvent receiving
vessel", page 28
6.5
Switching off the instrument
Precondition:
R
The instrument is shut down. See Chapter 6.4 "Shutting down the instrument",
page 24
u
Press the button
Stop
.
u
Disconnect the communication cable.
u
Disconnect the power cable.
6.6
Setting condenser temperature
Increase temperature
u
Press the button
UP
.
Decrease the temperature
u
Press the button
DOWN
.