BÜCHI Labortechnik AG
Cleaning and servicing | 7
Operation Manual Inert Loop B-295 SE
27/38
NOTICE! Check the filling level of the solvent
receiving vessel during the draining process.
If necessary empty it.
See Chapter 7.3 "Empty
the solvent receiving vessel", page 28
u
Drain the cleaning liquid through the conden-
sate drain valve and the process gas tube
connected to the connection marked
OUT
.
u
Repeat the cleaning procedure until the wash-
ing solution is free from particles or chemicals
u
Install the inlet tube to the connected spray
dryer. See Chapter 5.5 "Connecting the Inert
Loop B-295 SE with a spray drying system",
page 20
u
Place the end of the exhaust gas tube to a
fume hood.
u
Place the end of the outlet tube to a fume
hood.
u
Connect the power supply.
u
Press the button Switch off.
STOP
u
Establish a cold dry stream of drying gas to
dry the instrument.
u
Dry the instrument for 10 Minutes.
u
Remove the screw plug.
u
Install the oxygen sensor. See Chapter 8.2.2
"Installing the oxygen sensor", page 32