BTL VANQUISH ME - OPERATOR’S MANUAL I PAGE 35 OF 44
Touch Panel Calibration. If the touch screen buttons do not react when pressed, the touch
screen needs to be calibrated. A simple wizard controls the entire process. During calibration,
use the touch-pen (stylus) and follow the on-screen instructions. Calibration can be interrupted
at any time by pressing the esc button. To verify the touch screen adjustments, use the touch
panel function test function.
User Options. This submenu allows the setting and displaying of the following parameters
:
•
direction of movement of menu (standard / reverse)
•
tabs position (top / bottom)
•
speaker volume
Style of Operation. This submenu allows the setting and displaying of the following parameters:
•
set zero time – you can select whether after the end of therapy, the time values on the
screen shall remain zero, or whether the values of the last performed therapy will be
displayed
•
repeat sound for end – you can enable or disable the repeating audio signal which
indicates the end of therapy session
Unit Information. This submenu contains information about the device: the serial number, the
device type, the firmware version, etc. If the operation of the device is time restricted, this
submenu will indicate the date until when the device remains fully functional.
Unlock Code. Time restrictions imposed on the use of the BTL Vanquish ME system operation
can be removed by entering the unlock code. Unlock code can be obtained from your BTL
dealer or service technician.
Service Functions
•
Repair of files – Checks the file storage system. It repairs errors, deletes empty files, etc.
Use in case of a lack of memory space, if you have difficulties saving data, or if you
suspect data loss.
•
File system formatting – Use only if the repair of files did not help. When used, the entire
memory is re-formatted and all the user-data and the user setup are lost.
•
Delete accessories – Deletes all the previously installed accessory data. Use in case of
incorrectly installed applicator, or when the connected applicator cannot be detected
(question mark on the tab).
•
Default setting without loss of the user data – Restores all functions of the device to the
factory default settings. User-created data will not be deleted.
•
Unit log data export
6.6.2
SPECIFIC SETTINGS
Acoustic Signal Settings. Enable or disable acoustic signal in case of insufficient tuning.
Therapy Checkpoints Settings. Set therapy pause at a given interval (factory preset is 10
minutes). This feature can be disabled.
6.6.3
TREATMENT RECORDS
This submenu displays the list of the last performed treatments.