BTL VANQUISH ME - OPERATOR’S MANUAL I PAGE 32 OF 44
buttons to continue
or stop therapy
button to pause
therapy
6.4 START, INTERRUPTION AND END OF THERAPY
Set the desired parameters and start the therapy by pressing the start/stop key (5) on the unit
or the start button on the touch screen. Therapy can only be started if the therapy parameters
screen is displayed on the channel display and all parameters are set correctly. The BTL
Vanquish ME will warn the user about any discrepancies.
To interrupt the therapy, press the start/stop key (5) on the unit or the pause button on the
touch screen. While the therapy is interrupted (paused), you can modify the therapy time. To
restore the interrupted (paused) therapy, press the start/stop key (5) again or press the
continue button on the touch screen.
To stop the therapy, press the esc button while in the pause mode.
By default, the therapy ends after the set time elapses.
•
During the therapy, you can modify the output power using the select knob.
•
The patient can turn the unit off by pulling the Patient Safety Switch. For further details,
please refer to Chapter 4.4.
•
The BTL Vanquish ME control unit automatically monitors the system to help ensure the
safety of the patient and the device. Should any error occur, the output is shut off.
6.5 OUTPUT POWER
After the treatment parameters have been selected and therapy has been started, you may
increase or decrease the output power. To increase the output, turn the select knob (2) to the
right, to decrease the output, turn the select knob (2) to the left.
select
The default output power is 10 points. After any therapy interruption (pause), the therapy
re-starts at the output of 10 points.