MAINTENANCE AND SAFETY INSTRUCTIONS | PAGE 26 OF 38
4
MAINTENANCE AND SAFETY INSTRUCTIONS
A service inspection including the measuring of all parameters of the device and relevant recalibration must be performed at
intervals shorter than 30 months. The inspection and recalibration must be performed by an authorized BTL service department.
If the inspection is not done at 30-month intervals, the manufacturer does not guarantee the technical parameters or safe
operation of the product.
To keep the device clean, do not store or use it in extremely dusty environment for a long time. Do not immerse it in any liquid.
Before each use, checks that the device and its accessories (especially cables) are not mechanically or otherwise damaged. Do
not use the device if it is damaged!
Exterior cleaning of the device:
Use a soft cloth slightly moistened with water or with a 2% detergent solution to clean the exterior of the BTL-6000 SWT Topline
device and its parts. Never use cleaning agents containing alcohol, ammonia, benzine, thinners, etc. Never use abrasive
cleaning materials which will scratch the device's surfaces. No parts of the device require sterilization. Care should be given to
prevent water or other liquids from getting inside the device.
Cleaning and maintenance of accessories which come into contact with the patient:
Clean and disinfect after each client using approved cleaning agents. For example, Sekusept, Bacilol, and Incidur Spray can be
used. For the cables of accessories, use Incidur Spray and the alike. DO NOT USE SOLVENTS!!!
The exterior surface of the shock transmitter can be washed with warm water with its cap on. However, to completely remove
all of the contact gel from the shock transmitter, it will be necessary to unscrew the shock transmitter and clean it.
Fuse replacement:
The fuse is placed in the round black boxes on the rear panel. During replacement, check the correctness of the fuse being
inserted. This action should only be done by a person acquainted with this procedure!
Before replacement, make sure that the main power switch of the device is in the “0” position and the adapter is unplugged from
the unit. Turn the segment of the fuse box to the left using a flathead screwdriver or coin in the slot to remove the fuse. Insert a
new fuse and turn it to the right.
Do not use fuses other than those stated above the fuse box!
Servicing of the vessel for condensate:
The vessel for condensate on the rear side of the device shall be emptied and cleaned regularly. After 150,000 shot pulses the
device will prompt you automatically to pour out the condensate from the vessel. To release the vessel for condensate, screw it
out by turning it to the left; after pouring the condensate out screw the vessel back into position tightly.
Plugging the device into an electrical outlet:
The device is equipped with automatic voltage detection, so it can be used for voltages within the 100-240V.
Transport and Storage:
Keep the shipping container and all packaging materials. Transport the unit in original box to ensure maximum protection.
Unplug the main power cable and all accessory cables. Take care to avoid shocks or jarring movements to the device during
transport. This device should only be transported and stored under the conditions defined in the Technical Parameters section.