2.6.6
THERAPY PARAMETERS SCREEN – ERGONOMIC, STANDARD AND EXPERT MODE
This screen always appears before the start of therapy when the diag or prog buttons are pressed (see Diagram of the
therapy setting process). If only the most important therapy parameters are shown, then the ergonomic operation mode was
selected. If all therapy parameters are shown, then either the standard or the expert operation mode was selected. However, in
the expert mode all of these therapy parameters can be modified.
The differences between the operational modes can be seen in the following pictures:
+
Ergonomic mode
Standard mode
Expert mode
The operation mode can be set by pressing menu. For more details, see the Operation mode section. The expert operation
mode can be quickly entered from any mode by pressing the man button.
2.7
COURSE OF THERAPY
2.7.1
START, INTERRUPTION AND END OF THERAPY
Start the therapy by pressing the start button on the screen or by pressing the start/stop key on the front panel of the device.
The therapy can only start if the channel screen displays the therapy parameters and all those parameters are correctly set. The
device will warn of any possible discrepancies.
To interrupt therapy, press the pause button on the screen or the start/stop key on the panel of the device. To resume the
interrupted – paused – therapy press the start/stop key on the panel again or press the continual button. To stop, press esc.
During interrupted/paused therapy, the parameters of intensity and the total number of applied shocks can be modified by
turning the select knob.
The intensity can also be changed during therapy by turning the select knob to the right to increase the intensity or to the left to
decrease the intensity.
Shockwaves therapy can also be started or interrupted by using the button located on the shockwave applicator. The factory
default setting of the button is the AUTO mode. The device automatically recognizes the hold time of the button on the
applicator as well as the selected method of control. If the button is held for more than 1 second, then after its subsequent
release the therapy will stop (as in the continual mode, see the chapter APPLICATOR BUTTON MODE). During the time while
pressing the applicator button, generation of shockwaves will continue. If the applicator button is pressed for less than 1 second,
then after its release the device will be set to the ON/OFF option (see the chapter APPLICATOR BUTTON MODE).