INSTRUCTIONS FOR OPERATION | PAGE 24 OF 38 

2.11.3.14 Service Functions 

 

repair of files 

Checks the file storage system of the device, the system of saved information.  It will repair possible errors, delete empty files, 

etc. We recommend using this feature in the event of a lack of memory space, if the device rejects saving any data, or if you are 

in doubt that some data may have been lost. 

 

file system formatting 

Should be use only if the repair of damaged files feature did not help. Unfortunately, when this feature is used the entire memory 

is re-formatted and all the user-data and the user setup of the device will be lost. 

 

delete accessories 

Deletes  all  the  previously  installed  accessory  data.    It  is  to  be  used  when  accessories  were  incorrectly  installed  (with  bad 

“scattered"  picture  shown  on  the  tab),  when  the  connected  accessories  cannot  be  detected  (with  a  permanently  displayed 

question mark on the tab), or when there are other accessory related problems. 

 

default setting without loss of user data 

Restores  all  functions  of  the  device  to  the  factory  default  settings.  However,  all  user–data  related  to  clients,  programmed 

therapies, etc. will stay saved. 

2.11.4  SPECIFIC SETTINGS 

2.11.4.1  Applicator Button Mode 

This allows the selection of one of the three different types of modes for the applicator.The choices are: 

 

AUTO mode 

 

Continual mode 

 

ON/OFF mode 

In AUTO mode,  the  device  automatically  recognizes  the  hold  time  of  the  button  on  the  applicator  and  the  selected  way  of 

control it. If the button is pressed for more than 1 second, then after its subsequent release therapy will stop.  This will be the 

same in the continual mode (See below). During the time it takes to press the button , the generation of shockwaves will be in 

progress.  

If the button of the applicator is pressed for less than 1 second, then after its release the device will be set to the ON/OFF mode 

option  (see below). 

In  continual  mode,    the  generation  of  the  preset  shockwaves  is  accomplished  by  only  one  press  of  the  button  on  the 

shockwave applicator. To end the generation of shockwaves, it is necessary to press the button on the applicator. 

In  ON/OFF  mode,  it  is  necessary  to  hold  the  button  of  the  applicator  pressed  throughout  the  duration  of  the  application  of 

shockwaves. Releasing the button stops the generation of shockwaves. 

2.11.4.2  Applicator Kit Replacement Wizard 

If,  after  a  certain  amount  of  time,  the  applicator  stops  generating  shockwaves  correctly,  it  is  possible  to  change  wom-out  

applicator  parts.  Follow  the  instructions  detailed  in  the  Procedure  of  worn-out  tube  and  roller  replacement.  Install  this 

replaced kit according to the displayed instructions. 

Procedure for the applicator replacement kit: 

1. 

Connect the applicator to the device. 

2. 

Start the applicator kit replacement guide program.. 

3. 

Disconnect the original applicator from the device and complete the replacement of worn-out part following the step-by-step 

instructions shown in the guide  

4. 

After completion of the replacement re-connect the applicator back to the device. 

5. 

Complete the steps shown in the guide. 

6. 

The replaced kit has been successfully installed. 

Содержание 6000 SWT TOPLINE

Страница 1: ...GENERAL CHARACTERISTICS OF THE DEVICE PAGE 1 OF 38 BTL 6000 SWT TOPLINE USER S MANUAL v100AS08 01 2010EN...

Страница 2: ...ince we believe an ongoing relationship with our customers is critically important to our future product line While we would like you to start using your new equipment right away we encourage a thorou...

Страница 3: ...Therapy Parameters Via the diag Button 13 2 6 2 Setting Therapy Parameters Via the prog Button 13 2 6 3 Setting Therapy Parameters Manually User Setup Via the man Button 14 2 6 4 Setting the Number o...

Страница 4: ...2 11 4 1 Applicator Button Mode 24 2 11 4 2 Applicator Kit Replacement Wizard 24 3 LIST OF STANDARD AND OPTIONAL ACCESSORIES 25 4 MAINTENANCE AND SAFETY INSTRUCTIONS 26 4 1 1 Shock Transmitter Replace...

Страница 5: ...Topline is the predefined programs stored in the memory of the main unit Based on detailed research and practical use of the device the well organized predefined programs will provide recommendations...

Страница 6: ...positive pressure pulse Its diffusion is only limited by the actual tissue absorption and eventual reflections on acoustic non homogeneities A shockwave is defined as a pressure pulse with these char...

Страница 7: ...e tension Analgesic effect o Destruction of afferent nerves and nerve receptors o CNS stimulants sensed as pain are also inflammation transmitter substances o Regression of pain caused by local ischem...

Страница 8: ...ade 1 Arthrosis Acupuncture Pain trigger points or painful points in muscles 1 8 CONTRA INDICATIONS FOR SHOCKWAVE TREATMENT Application to certain tissues The eyes and the surrounding area the myocard...

Страница 9: ...t and stop therapy 6 ON OFF switch back lit in blue when the control unit is on 7 USB port in the space of the device s grip for use only in compliance with IEC 60950 1 The USB port serves only for se...

Страница 10: ...00 SWT TOPLINE 14 connector for shockwave applicator 15 control unit mains fuse 16 connector for power cable 17 power on off switch 18 vessel for collecting condensed water 19 type label contains type...

Страница 11: ...estions please contact an authorized service of BTL devices Procedure 1 First connect the device in mains by means of the supplied power supply adapter which you will connect to the connector on the r...

Страница 12: ...ative purposes but others can be pressed and activated These basic elements include 3D buttons which can be pressed to change the indicated values Informative text Channels tabs to switch between the...

Страница 13: ...AMETERS VIA THE DIAG BUTTON A list of diagnosis will display after pressing the diag button Each tab channel has its own assigned list of diagnosis To find a diagnosis quickly press the button for the...

Страница 14: ...dual buttons will open an individual menu and pop up boxes for the settings Most of the pop ups are accompanied with illustrative pictures and or symbols 2 6 4 SETTING THE NUMBER OF SHOCKS The total n...

Страница 15: ...cks 2 6 5 SETTING THERAPY INTENSITY The intensity power of the shockwave therapy can be set on the therapy parameters screen even during the course of therapy After pressing the intensity button it is...

Страница 16: ...ameters are correctly set The device will warn of any possible discrepancies To interrupt therapy press the pause button on the screen or the start stop key on the panel of the device To resume the in...

Страница 17: ...the treated tissue during therapy Minimizes the transmission of reverse shocks to the hand of the therapist due to the built in air damper An integral part of the applicator is the ergonomic hand rest...

Страница 18: ...r These examples illustrate the application of shockwaves in different areas and several of ways of holding the applicator Painful shoulder Trigger points Tibial Edge Syndrome Calcar Calcanei Plantar...

Страница 19: ...ed in both lists below the line After entering this data it is possible to assign a unique name to that particular therapy program The name of the therapy program will consist of the letter W for shoc...

Страница 20: ...e of the recently used therapies on the selected tab After pressing the load button the therapy can be restarted and its parameters viewed 2 11 UNIT SETTINGS THE MENU BUTTON After pressing menu it is...

Страница 21: ...d shocks etc 2 11 2 ENCYCLOPAEDIA The encyclopaedia provides information about diagnoses possible therapies and examples of how to use the device A hard copy of the encyclopaedia is always supplied wi...

Страница 22: ...f standard sounds from the manufacturer for example the signaling the therapy operations All audio tones can be muted no sound or modified as required Allows the editing of individual audio schemes th...

Страница 23: ...rences between the modes can be best seen on the therapy settings screen which will always appears before starting of therapy 2 11 3 9 Touch Panel Calibration If the buttons on the touch screen do not...

Страница 24: ...e selected way of control it If the button is pressed for more than 1 second then after its subsequent release therapy will stop This will be the same in the continual mode See below During the time i...

Страница 25: ...rom BTL For more detailed information on individual accessories see the enclosed leaflet Standard accessories 1x shockwave applicator with multi focus shock transmitter 15mm 1x replaceable multi focus...

Страница 26: ...surface of the shock transmitter can be washed with warm water with its cap on However to completely remove all of the contact gel from the shock transmitter it will be necessary to unscrew the shock...

Страница 27: ...ti focus shock transmitter 15mm 1x replaceable multi focus shock transmitter 9mm 1x replaceable focusing shock transmitter 15mm Replacement procedure 1 Unscrew the screw cap of the shock transmitter b...

Страница 28: ...parts kit included in the BTL 6000 SWT Topline package Do not use damaged applicators There is a risk of injury to the operating staff or the client The spare parts kit includes 1x exchangeable kit 1x...

Страница 29: ...STRUCTIONS PAGE 29 OF 38 3 Use the key to unscrew the screw cap of the cartridge See the pictures below 4 Remove the worn out part from the cartridge from the applicator 5 Remove the plastic cover fro...

Страница 30: ...little bit inside the applicator case The replaced part must move freely 7 After inserting the replaced part firmly screw the screw cap of the cartridge back in place using the key 8 When replacing th...

Страница 31: ...the Technical Parameters section It must not be used in an environment where there is a danger of explosion or of water penetrating the device The device cannot be in connection with flammable anaesth...

Страница 32: ...paired or replaced by the user Do not remove the cover from the control unit All repairs must be done by an authorized BTL service department Do not disconnect an applicator during therapy 4 3 USED SY...

Страница 33: ...127 2 Power switch according to IEC 60601 1 On the back of device positions 0 and I To disconnect from the mains unplug the male plug of the power supply adapter from the mains socket outlet Design We...

Страница 34: ...529 IP20 Class according to IEC 60601 1 II Class according to MDD 93 42 EEC Ilb Type of operation permanent Operating conditions Ambient temperature 10 C to 40 C Relative humidity 30 to 75 Atmospheric...

Страница 35: ...pline is suitable for use in the specified electromagnetic environment The purchaser or user BTL 6000 SWT Topline should assure that it is used in an electromagnetic environment as described below Imm...

Страница 36: ...e 2 note These are guidelines Actual conditions may vary Recommended separation distance BTL 6000 SWT Topline is intended to be used in electromagnetic environment with controlled RF disturbances The...

Страница 37: ...Testing and measurement techniques Electrical fast transient burst immunity test IEC 61000 4 4 am1 ed2 0 2009 EN ISO 61000 4 4 2005 Electromagnetic compatibility EMC Part 4 Testing and measurement tec...

Страница 38: ...ransferred by any means including electronic mechanical and photographic without the prior written approval from BTL Industries Limited BTL Industries Limited operates on a policy of continuous develo...

Отзывы: