INSTRUCTIONS FOR OPERATION | PAGE 24 OF 38
2.11.3.14 Service Functions
•
repair of files
Checks the file storage system of the device, the system of saved information. It will repair possible errors, delete empty files,
etc. We recommend using this feature in the event of a lack of memory space, if the device rejects saving any data, or if you are
in doubt that some data may have been lost.
•
file system formatting
Should be use only if the repair of damaged files feature did not help. Unfortunately, when this feature is used the entire memory
is re-formatted and all the user-data and the user setup of the device will be lost.
•
delete accessories
Deletes all the previously installed accessory data. It is to be used when accessories were incorrectly installed (with bad
“scattered" picture shown on the tab), when the connected accessories cannot be detected (with a permanently displayed
question mark on the tab), or when there are other accessory related problems.
•
default setting without loss of user data
Restores all functions of the device to the factory default settings. However, all user–data related to clients, programmed
therapies, etc. will stay saved.
2.11.4 SPECIFIC SETTINGS
2.11.4.1 Applicator Button Mode
This allows the selection of one of the three different types of modes for the applicator.The choices are:
•
AUTO mode
•
Continual mode
•
ON/OFF mode
In AUTO mode, the device automatically recognizes the hold time of the button on the applicator and the selected way of
control it. If the button is pressed for more than 1 second, then after its subsequent release therapy will stop. This will be the
same in the continual mode (See below). During the time it takes to press the button , the generation of shockwaves will be in
progress.
If the button of the applicator is pressed for less than 1 second, then after its release the device will be set to the ON/OFF mode
option (see below).
In continual mode, the generation of the preset shockwaves is accomplished by only one press of the button on the
shockwave applicator. To end the generation of shockwaves, it is necessary to press the button on the applicator.
In ON/OFF mode, it is necessary to hold the button of the applicator pressed throughout the duration of the application of
shockwaves. Releasing the button stops the generation of shockwaves.
2.11.4.2 Applicator Kit Replacement Wizard
If, after a certain amount of time, the applicator stops generating shockwaves correctly, it is possible to change wom-out
applicator parts. Follow the instructions detailed in the Procedure of worn-out tube and roller replacement. Install this
replaced kit according to the displayed instructions.
Procedure for the applicator replacement kit:
1.
Connect the applicator to the device.
2.
Start the applicator kit replacement guide program..
3.
Disconnect the original applicator from the device and complete the replacement of worn-out part following the step-by-step
instructions shown in the guide
4.
After completion of the replacement re-connect the applicator back to the device.
5.
Complete the steps shown in the guide.
6.
The replaced kit has been successfully installed.