4.6 - Electronics box display menu
M2007/3en
TCT-2301 Operating instructions
69
©BTG 2004
4.6.2
Menu structure
!
After 15 seconds, the transmitter always returns to the result
display regardless of where it was in the menu structure.
Press the arrow buttons to step between menu items.
Start up display
TCT-2301
Ver x.x
Shows transmitter model TCT-2301 and
software version.
Result display 1
1.30%
TCONS
Result display, shows consistency.
Result display 2
12.53mA TCONS
Result display, shows consistency output signal.
Range setting
RANGE 1
Current measuring range, can be changed to feed in
values for other ranges (range 1, 2, 3 and 4).
Span setting
SPAN
Measuring span, low limit = 4mA high limit = 20mA.
Offset adjustment
OFFSET
Consistency offset if required.
Damping setting
DAMPING
Damping on output signal.
Calibration setting
CALIBRATION
Calibration constants for the transmitter.
Cleaning cycle
CLEANING
The cleaning system is an option for the TCT.
All time settings to control the system are made here.
Language
selection
LANGUAGE
en
Language setting, in this case English.