67
Bruker Optik GmbH
HYPERION User Manual
Operation 4
5
For verifying the currently detected IR
signal intensity, compare the amplitude
value displayed in OPUS (see image in
step 4) with the amplitude value for the
transmission mode stated in the supplied
OVP test protocol (see image in step 5).
The supplied test protocol
Hyperion OVP
Supplementary Sheet
documents the
result of a factory-performed signal check
using your analysis system. The test has
been performed with all optics compo-
nents being optimally adjusted. You will
find the test protocol in the folder supply
with the microscope.
6
If there is not any IR signal detected or if the amplitude value displayed in OPUS deviates
significantly from the amplitude value of the supplied OVP test protocol, see section 7.2
for possible causes and troubleshooting solutions.
Note: A detected amplitude value above 32.000 is an indication of an oversaturated
detector. In this case you first have to reduced the signal intensity until the detector is no
longer oversaturated. For information about how to reduce the signal intensity, see
section 4.8.
Amplitudenwert im
Transmissionsmodus
Содержание HYPERION
Страница 1: ...HYPERION User Manual I 24319 ...
Страница 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Страница 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Страница 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Страница 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Страница 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Страница 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Страница 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...
Страница 180: ...HYPERION User Manual Bruker Optik GmbH ...