132
HYPERION User Manual
Bruker Optik GmbH
Repair and Maintenance 6
7.
Push the adapter carefully over the connection nozzle of detector dewar and fasten
the coupling nut (A in fig. 6.3) hand-tight while holding the adapter and the detector
as shown in fig. 6.2b. (A hand-tight tightening of the coupling nut is sufficient.)
➣
Hold the adapter and the detector always as shown in fig. 6.2b when you have to
carry out the following tasks: opening and closing the evacuation valve by pushing
or pulling the adapter knob (step 8, 12, 14 and 17), screwing the threaded adapter
rod in or out of the connection thread of the dewar evacuation valve (step 11 and
15) and loosening the coupling nut (step 18).
8.
Push the adapter knob (G in fig. 6.3) in the closed position until the threaded rod (D
in fig. 6.3) of the adapter is in contact with the sealing plug of the dewar evacuation
valve.
9.
Before you begin to evacuate the detector dewar, check the connections for leak
tightness by evacuating the section between vacuum pump and detector at first. To
do this, open the shut-off valve. If a vacuum of 10
-4
mbar is generated within a few
minutes it is an indication of the leak tightness of this section.
10. Close the shut-off valve again.
11. Screw the threaded rod of the adapter (D in fig. 6.3) in the connection thread of
dewar evacuation valve by turning the adapter knob (G in fig. 6.3) clockwise; 2 to 3
rotations are sufficient.
Attention: In case of more than 2 or 3 knob rotations
there is a risk that the threaded connection becomes inseparable! That means
the threaded rod of the adapter cannot be screwed out of the connection
thread of the dewar evacuation valve again.
12. Pull the knob (G in fig. 6.3) to the open position in order to open the dewar evacua-
tion valve.
13. Begin to evacuate the detector dewar by opening the shut-off valve.
➣
We recommend an evacuation time of at least 3 days to allow for generating an opti-
mal vacuum inside the dewar. The final pressure in the detector dewar should be
less than 10
-5
mbar.
Figure 6.2:
b) How to hold the MCT detector when oping or
closing the dewar evacuation valve
Figure 6.2:
a) MCT detector
Connection nozzle
(without cap)
Содержание HYPERION
Страница 1: ...HYPERION User Manual I 24319 ...
Страница 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Страница 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Страница 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Страница 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Страница 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Страница 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Страница 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...
Страница 180: ...HYPERION User Manual Bruker Optik GmbH ...