To reinstall the MCT detector, perform the above steps in reverse order:
1.
Insert the MCT detector carefully into the detector compartment.
2.
Fasten the detector base plate.
3.
Reconnect all cables to the MCT detector.
4.
Reinstall the detector compartment housing.
5.
Switch on the spectrometer.
Step 6
Loosen the two mounting screws at
the base plate of MCT detector
using an Allen wrench (3 mm).
Step 7
Remove the MCT detector carefully
from the detector compartment.
Caution: Be careful not to dam-
age the mirrors!
Mounting screws