
137
Bruker Optik GmbH
HYPERION User Manual
Repair and Maintenance 6
Procedure
For replacing the crystal assembly, the ATR objective needs to be removed from the
microscope.(See section 4.12.3.) Put the ATR objective on an even surface with the
crystal pointing upwards.
Required tools:
• small flat tip screwdriver
• tweezers ((Note: They are in the tool box supplied with the microscope.))
• hex wrench (Note: This hex wrench is included in the delivery scope of the ATR
crystal.)
1x set of mounting hardware
Crystal assembly
Figure 6.5:
Replacement ATR crystal - Delivery scope
Step 1
Loosen and remove the three slotted-
head screws at the outer ring using a
small flat tip screwdriver.
Slotted-head screws at the outer ring
Содержание HYPERION
Страница 1: ...HYPERION User Manual I 24319 ...
Страница 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Страница 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Страница 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Страница 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Страница 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Страница 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Страница 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...
Страница 180: ...HYPERION User Manual Bruker Optik GmbH ...