119
Bruker Optik GmbH
HYPERION User Manual
Options and Accessories 5
5.3.3.2
Safety notes
The sample holder can be heated up to 180°C. So during operation, the sample holder
can become very hot.
➣
Attention:
Depending on the property of the sample, it can dramatically change its
physical appearance (e.g. melt) or even ignite. For this reason, determine in
advance how the sample in question will behave at the intended temperature.
5.3.3.3
Controlling the temperature
A temperature controller is included in the delivery scope of the heatable sample holder.
(See fig. 5.4b.) The desired temperature can be set as follows:
The temperature is kept constant by means of a temperature sensor and a temperature
controller.
i
Temperature information is stored together with the spectra if the temperature has been
controlled by the OPUS software. Measurements during temperature ramps are only
possible with the OPUS measurement function
Protein Dynamics
(optional software
package OPUS/PRO).
Figure 5.4: a) Heatable sample holder
(placed on the microscope
stage)
Figure 5.4: b) Temperature controller
CAUTION
Risk of burn injuries caused by touching the heatable sample holder
➣
Do not touch the hot sample holder.
➣
Wait until the sample holder and the sample have cooled down before touching
the holder or removing the sample.
computer-controlled by means of the
OPUS measurement function
Tempera-
ture Control
☞
For detailed information about this
OPUS function, refer to the OPUS
Reference Manual.
computer-controlled by means of the
OPUS measurement function
Protein
Dynamic
☞
For detailed information about this
OPUS function, refer to the OPUS/
PRO Manual.
manually by means of the temperature
controller (fig. 5.4b)
☞
Refer to the accompanying manual.
Содержание HYPERION
Страница 1: ...HYPERION User Manual I 24319 ...
Страница 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Страница 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Страница 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Страница 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Страница 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Страница 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Страница 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...
Страница 180: ...HYPERION User Manual Bruker Optik GmbH ...