78
ZTKS0178 / Z31970 / Rev.: 03
Technical Data MS 750‘89 Ascend
Temperature Sensors
The temperature sensors (PT 100 and IBT) will be used to monitor the temperature of
the magnets during cooling and warming up of the magnet system.
PT 100
i
Measure the resistance with a maximum current of 1 mA.
Temperature
Unit
Resistance
Unit
Room Temperature
293
K
107.8
Ω
273
K
100.0
Ω
250
K
91.0
Ω
200
K
71.1
Ω
150
K
50.9
Ω
100
K
30.0
Ω
Liquid Nitrogen
77
K
20.1
Ω
Table A.7:
Characteristic Values PT 100
Figure A.7:
Characteristic Curve PT 100
Содержание Ascend 750'89
Страница 1: ...Version Innovation with Integrity NMR Spectroscopy 750 89 Ascend 03 User Manual ...
Страница 4: ...4 ZTKS0178 Z31970 Rev 03 ...
Страница 7: ...7 ZTKS0178 Z31970 Rev 03 ...
Страница 8: ...8 ZTKS0178 Z31970 Rev 03 π ...
Страница 10: ...10 ZTKS0178 Z31970 Rev 03 Contact ...
Страница 34: ...34 ZTKS0178 Z31970 Rev 03 Transportation ...
Страница 35: ...35 ZTKS0178 Z31970 Rev 03 4 Assembling 4 1 Safety Approved Persons Bruker Service only ...
Страница 36: ...36 ZTKS0178 Z31970 Rev 03 Assembling ...
Страница 57: ...57 ZTKS0178 Z31970 Rev 03 8 Disassembling 8 1 Safety Approved Persons Bruker Service only ...
Страница 58: ...58 ZTKS0178 Z31970 Rev 03 Disassembling ...
Страница 59: ...59 ZTKS0178 Z31970 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 60: ...60 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 62: ...62 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 64: ...64 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 66: ...66 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 68: ...68 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 75: ...75 ZTKS0178 Z31970 Rev 03 Technical Data MS 750 89 Ascend Helium Level Graph Figure A 4 Helium Level Graph ...
Страница 83: ...83 ZTKS0178 Z31970 Rev 03 Technical Data MS 750 89 Ascend Current Lead Figure A 11 Current Lead 55 pins ...
Страница 94: ...94 ZTKS0178 Z31970 Rev 03 Technical Data MS 750 89 Ascend This page is intentionally left blank ...