50
ZTKS0178 / Z31970 / Rev.: 03
Troubleshooting
6.2.6
During Standard Operation
Continued on next page
Indicator
Possible reason
Solution
By
The helium boil off
decreases to zero.
The atmospheric
pressure is
increasing.
Usual behavior.
Watch helium boil off daily.
Approved
Customer
Personnel
The helium flow
system is covered
with ice.
Contact Bruker Service imme-
diately! Do not try to remove
ice of the helium flow system!
Approved
Customer
Personnel
WARNING:
Cryogenic Agents
Quench
The helium flow
system or the
suspension tubes
are blocked with ice.
Blow in warm helium gas
carefully through an
applicable tube. Do not insert
it more than 600 mm from the
top of the helium turrets.
Bruker Service
The helium boil off is
too high.
The Helium Level
Sensor is
permanently on
(service mode) or
used often.
Switch off Helium Level
Sensor. Reduce frequency of
helium level measurement
(during measuring of the
helium level an amount of
helium boils off due to the
heat input of the Helium Level
Sensor;
for reporting the helium boil
off one weekly measurement
is sufficient).
Approved
Customer
Personnel
The atmospheric
pressure is
decreasing.
Usual behavior.
Watch helium boil off daily.
Approved
Customer
Personnel
If the problem is still not
solved, contact
Bruker Service.
Approved
Customer
Personnel
Содержание Ascend 750'89
Страница 1: ...Version Innovation with Integrity NMR Spectroscopy 750 89 Ascend 03 User Manual ...
Страница 4: ...4 ZTKS0178 Z31970 Rev 03 ...
Страница 7: ...7 ZTKS0178 Z31970 Rev 03 ...
Страница 8: ...8 ZTKS0178 Z31970 Rev 03 π ...
Страница 10: ...10 ZTKS0178 Z31970 Rev 03 Contact ...
Страница 34: ...34 ZTKS0178 Z31970 Rev 03 Transportation ...
Страница 35: ...35 ZTKS0178 Z31970 Rev 03 4 Assembling 4 1 Safety Approved Persons Bruker Service only ...
Страница 36: ...36 ZTKS0178 Z31970 Rev 03 Assembling ...
Страница 57: ...57 ZTKS0178 Z31970 Rev 03 8 Disassembling 8 1 Safety Approved Persons Bruker Service only ...
Страница 58: ...58 ZTKS0178 Z31970 Rev 03 Disassembling ...
Страница 59: ...59 ZTKS0178 Z31970 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 60: ...60 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 62: ...62 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 64: ...64 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 66: ...66 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 68: ...68 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 75: ...75 ZTKS0178 Z31970 Rev 03 Technical Data MS 750 89 Ascend Helium Level Graph Figure A 4 Helium Level Graph ...
Страница 83: ...83 ZTKS0178 Z31970 Rev 03 Technical Data MS 750 89 Ascend Current Lead Figure A 11 Current Lead 55 pins ...
Страница 94: ...94 ZTKS0178 Z31970 Rev 03 Technical Data MS 750 89 Ascend This page is intentionally left blank ...