48
ZTKS0178 / Z31970 / Rev.: 03
Troubleshooting
Continued from page before
6.2.5
During Operation of the Magnet Stand
In case of doubt contact Bruker Service and refer to the manual of the Magnet Stand.
Continued on next page
Indicator
Possible reason
Solution
By
Magnet system does
not reach
specification.
Magnetic material
inside RT bore tube.
Carefully clean the RT bore
tube.
Bruker Service
Large ferromagnetic
parts near the
magnet system.
1.
Keep the maximum pos-
sible distance between
the magnet system and
ferromagnetic parts.
2.
Repeat shimming.
Bruker Service
Indicator
Possible reason
Solution
By
The NMR spectrum
shows massive dis-
turbances.
Pneumatic controller
is in DOWN position.
Switch pneumatic controller to
UP position.
Approved
Customer
Personnel
Magnet system has
direct mechanical
contact with the floor
via accessories or
cables.
Identify and eliminate contact
point. Arrange cables in loose
S- or U-shapes.
Approved
Customer
Personnel
If the problem is still not
solved, contact
Bruker Service.
Approved
Customer
Personnel
Magnet system has
physical contact to
the magnet stand.
Check leveling; adjust if
necessary.
Bruker Service
Piston of the isolator
is not centric or
touches its casing.
Align magnet stand.
Bruker Service
T-safety bracket
touches the pillar.
Align magnet stand.
Bruker Service
Floor vibrations in
vertical direction.
Replace elastomeric isolators
with air damped isolators.
Bruker Service
Floor vibrations in
horizontal and
vertical direction.
Replace air damped isolators
with air piston isolators.
Bruker Service
Содержание Ascend 750'89
Страница 1: ...Version Innovation with Integrity NMR Spectroscopy 750 89 Ascend 03 User Manual ...
Страница 4: ...4 ZTKS0178 Z31970 Rev 03 ...
Страница 7: ...7 ZTKS0178 Z31970 Rev 03 ...
Страница 8: ...8 ZTKS0178 Z31970 Rev 03 π ...
Страница 10: ...10 ZTKS0178 Z31970 Rev 03 Contact ...
Страница 34: ...34 ZTKS0178 Z31970 Rev 03 Transportation ...
Страница 35: ...35 ZTKS0178 Z31970 Rev 03 4 Assembling 4 1 Safety Approved Persons Bruker Service only ...
Страница 36: ...36 ZTKS0178 Z31970 Rev 03 Assembling ...
Страница 57: ...57 ZTKS0178 Z31970 Rev 03 8 Disassembling 8 1 Safety Approved Persons Bruker Service only ...
Страница 58: ...58 ZTKS0178 Z31970 Rev 03 Disassembling ...
Страница 59: ...59 ZTKS0178 Z31970 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 60: ...60 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 62: ...62 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 64: ...64 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 66: ...66 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 68: ...68 ZTKS0178 Z31970 Rev 03 Appendix ...
Страница 75: ...75 ZTKS0178 Z31970 Rev 03 Technical Data MS 750 89 Ascend Helium Level Graph Figure A 4 Helium Level Graph ...
Страница 83: ...83 ZTKS0178 Z31970 Rev 03 Technical Data MS 750 89 Ascend Current Lead Figure A 11 Current Lead 55 pins ...
Страница 94: ...94 ZTKS0178 Z31970 Rev 03 Technical Data MS 750 89 Ascend This page is intentionally left blank ...